SECTION 2 PATIENT AND OPERATING ROOM SAFETY
Gyrus ACMI PlasmaKinetic SuperPulse Generator (Endourology)
USER MANUAL
Part Number: 144020-LB
Carefully insert and withdraw active instruments from sheaths and working
channels to avoid the possibility of damage to the devices and/or injury to
the patient.
Only activate during intermittent or preferably continuous flow irrigation to
ensure good visualisation and cooling of the instrument tip between
activations.
The vaporisation process produces bubbles. Activation of the instrument tip
within a bubble pocket may cause product damage.
Only use normal saline irrigation solution. Do not activate when not in
contact with tissue, or excessive heating of the irrigation medium may result.
Tissue contact with the return instrument whilst the active instrument is
surrounded by normal saline during activation may result in an
electrosurgical effect occurring at the return instrument.
CAUTION
Proper use of the system relies on tissue contact with the active tip of the
instrument only. Do not bury the tip in tissue beyond the junction of the active
instrument and the ceramic insulator as product damage may result during use.
WARNING
Olympus TURis Mode
Incorrect assembly and connection of the Olympus Resectoscopic system and
electrode may result in no RF surgical output.
IMPORTANT
TURis compatibility is only available with SuperPulse software version V3.0
software and later.
2F. After Surgery
WARNING
Electric Shock Hazard. Always unplug the generator before cleaning.
CAUTION
Do not reuse or resterilize accessories labelled “disposable” or “single use only.”
2G EMC Classification
The SuperPulse System has been manufactured and tested to the following requirements:
Group 2 Class A as per IEC60601-1-2 (2001)
EMC PRECAUTIONS
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information in this document.
EMC WARNINGS
• The generator should not be used adjacent to or stacked with other electrical equipment. If
adjacent or stacked use is necessary both the generator and other equipment should be observed
to verify normal operation in the configuration in which it will be used.
• The EMC classification of the SuperPulse system (class A) is suitable for use on dedicated supply
systems not connected to the public mains network, such as hospitals.
Page 2-6
SECTION 2
PATI
ENT AND OPERATING ROOM SAFETY