![Gymna Pulson 330 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/gymna/pulson-330/pulson-330_service-manual_2280458009.webp)
6. 1/3 MHz head selection key.
Is a key to switch from the 1MHz to the 3MHz head and back.
For a good understanding, some important notes have to be taken into
consideration:
The right side panel has 2 US-output connections, CON2 and CON3.
The pulson 330 is designed in such a way that it always detects
which head is
connected to which output gate.
In other words, both heads can be connected to both outputs, and it
is of no
importance whether one ore two heads are connected.
The following examples may illustrate this:
If both heads are connected and you switch on the instrument, it
will always
select the 1MHz head. If only the 3MHz head is connected, there
is no other choice,
so this one will be selected.
If you want to switch from 1MHz to 3MHz, there are two
possibilities:
The 3MHz is connected.
Press the 1/3 MHz and the switch is made.
The 3MHz is not connected.
The instrument doesnít find the other head. The 3MHz led
flashes for 10 sec.
after which the 1MHz is selected again.
If only the 1 MHz is connected, and you pull out this head, both
leds start flashing
for about 10 sec., after which they extinguish.
If two heads are connected and you pull ot the 1MHz while using it
for a treatment,
the instrument will immediately switch off the US-power and
select the 3MHz head.
7. Timer down switch.
8. Timer up switch.
The treatment time can be changed in steps of a min. if time > 1
min. If the treatment
time < 1 min. the timer can be changed in steps of a sec.
Page 9
Содержание Pulson 330
Страница 12: ......
Страница 16: ......
Страница 18: ......
Страница 22: ......
Страница 27: ......
Страница 29: ......
Страница 37: ...21 The ultrasound heads Page 37...
Страница 38: ...Page 38...
Страница 41: ...PARTS RETURN TAG Page 41...