16. Measuring points.
Anode D22: 5 V stabilised.
Kathode D21: - 5 V stabilised.
Following values can only be measured when the US - box is opened.
L12: 32 VDC.
Pin 1 CON 1/2: + 15 V stabilised.
R35: + 15 V +/- 1%.
Pin 14 of U15: + 5 V DC.
16.1 Cable fraction detection
Signal measured on efficiency resistance :
Ground is housing of the plug.
Cab
S2
S1
ch2: rms = 1.58 V
MTB5.00ms
2
Switch the unit on without starting the treatment.
Connect the head which has produced the code Cab.
The signal S2 is a 1.5 VDC tension, which is displayed under normal
functioning.
In case of a cable fraction, signal S1 appears.
Page 31