PSW Series User Manual
52
4.
Use the current knob to set the F-03
setting.
Set F-03 to 0 (CV High Speed
Priority) or 2 (CV Slew Rate
Priority).
Current
F-03
0 = CV High Speed Priority
2 = CV Slew Rate Priority
5.
Press the Voltage knob to save the
configuration setting. ConF will be
displayed when successful.
Voltage
6.
If CV Slew Rate Priority was chosen as the
operating mode, repeat steps 3~5 to set F-04
(Rising Voltage Slew Rate) and the F-05 (Falling
Voltage Slew Rate) and save.
F-04 / F-05
0.1V/s~60V/s (PSW 30-XX)
0.1V/s~160V/s (PSW 80-XX)
0.1V/s~320V/s (PSW160-XX)
7.
Press the Function key again to exit
the configuration settings. The
function key light will turn off.
Function
8.
Use the Current knob to set the
current limit (crossover point).
Current