PEL-3211(H) User Manual
30
Caution
Ensure that the maximum input voltage is not
exceeded. The maximum input voltage is 150(800)
volts.
Warning
Do not touch any of the input terminals when the unit
is on.
Warning
Connecting the input terminals to the wrong polarity
can damage the DUT or the booster packs/master
unit.
Load Connection
1.
Make sure the master unit and booster packs
are turned off.
2.
Make sure the DUT is turned off.
3.
Connect the load units to the DUT.
Ensure the wire gauge is sufficient to handle the
increase in current.
Connect the
rear
negative terminals on the
booster(s) and master unit to the negative
potential of the DUT.
Connect the
rear
positive terminals on the
booster(s) and master unit to the positive
potential on the DUT.
If possible use the shortest cable length
possible.
If possible, twist the load cables together to
reduce load line inductance.
Try to use cables of equal length.
Do not coil or loop the load cables.