7
Example:
ONLINE step 1(power on only)
PC send command
COMU?
<
^END^M
> or
<^J^M>
ASCII CODE =
43 4F 4D 55
3F 0A 0D (Hex format)
LCR Meter Response :
COMU
:
ON..
<
^END
>
ASCII CODE =
43 4F 4D 55
3A
4F 4E 2E
2E
0A (Hex format)
ONLINE step 2(power on only, LCR METER WILL SHOW RS232 ONLINE)
PC send command
COMU
:OVER
<
^END^M
> or
<^J^M>
ASCII CODE =
43 4F 4D 55
3A
4F 56 45 52
0A 0D (Hex format)
LCR Meter Response :
COMU
:
OVER
<
^END
ASCII CODE =
43 4F 4D 55
3A
4F 56 45
52
0A (Hex format)
Set Speed
PC send command
MAIN
:
SPEE
:
FAST
<
^END^M
> or
<^J^M>
ASCII CODE =
4D 41 49 4E
3A
55 50 45 45
3A
46 41 53 54
0A 0D (Hex format)
LCR Meter Response :
MAIN
:
SPEE
:
FAST
<
^END
>
ASCII CODE =
4D 41 49 4E
3A
53 50 45 45
3A
46 41 53 54
0A (Hex format)
Set Frequency
PC send command
MAIN
:
FREQ
1.00000
<
^END^M
> or
<^J^M>
ASCII CODE =
4D 41 49 4E
3A
46 52 45 51
20
31 2E 30 30 30 30 30
0A 0D (Hex format)
LCR Meter Response:
MAIN
:
FREQ
1.00000
<
^END
>
ASCII CODE =
4D 41 49 4E
3A
46 52 45 51
20
31 2E 30 30 30 30 30
0A (Hex format)