GOM-804 & GOM-805
使用手册
1
33
SYSTem:VOLTage:PROTect Set / Query
Set
Query
Description
Sets or returns the HVP function state.
Syntax
Query Syntax
SYSTem:VOLTage:PROTect <NR1> | {OFF | ON}
SYSTem:VOLTage:PROTect?
Parameter/
Return parameter
<NR1> 0:OFF.
1:ON
OFF Turn the HVP function off
ON Turn the HVP function on.
Example
SYST:VOLT:PROT OFF
Sets the HVP function off.
SYSTem:ERRor
Query
Description
Returns the current system error, if any.
Query Syntax
SYSTem:ERRor?
Return parameter
<String>
Error number,”Error message”
Example
SYST:ERR?
>0,”No error”.
Indicates that there is no error message.
Error Message List
0,”No error”
1,”Command error”
4,"Data out of range"
SYSTem:HANDler
Set
Query
Description
Sets or returns the handler state.
Syntax
Query Syntax
SYSTem:HANDler {CLEAR | HOLD}
SYSTem:HANDler?
Parameter/
Return parameter
Clear
It clears the last result before
executing measurement.
HOLD
It holds the test result and changes
when a different result appears.
Содержание GOM-804
Страница 1: ...直流微欧姆计 GOM 804 GOM 805 使用手册 固纬料号 ISO 9001 认证企业 ...
Страница 21: ...GOM 804 GOM 805 使用手册 2 1 设置 倾斜站立 倾斜 将支脚向前扳 如下所示 直立 将支脚向后扳 如下所示 ...
Страница 38: ...GOM 804 GOM 805 使用手册 38 Internal trigger source ...
Страница 57: ...GOM 804 GOM 805 使用手册 57 GUARD SOURCE SENSE SENSE SOURCE shielding ...
Страница 96: ...GOM 804 GOM 805 使用手册 9 6 功能检测完成 ...
Страница 137: ...GOM 804 GOM 805 使用手册 1 37 温度指令 ...