GW Instek GFC-8131H Скачать руководство пользователя страница 5

  FREQUENCY COUNTER

USER MANUAL

8

z

 

Front Panel

       

Fig. 1 Front panel

  FREQUENCY COUNTER

USER MANUAL

9

5. APPLICATION

(1)

 

Signal Measurements

Frequency range between 0.01Hz and 120MHz, press FREQ A switch,

then connect input signal to the channel A input BNC.

Frequency range between 50MHz and 1.3(2.7) GHz, press FREQ B

switch, then connect input signal to the channel B input BNC.

Press “PRID” switch to select period mode of operation for signal on

input A.

(2)

 

Gate Time Settings

The instruments features continuously adjustable gate time selection

from 10ms to 10s or one period of input, depending on whichever is

longer. The GATE TIME adjustment affects the sampling rate and the

resolution of the reading.

Turn the knob counterclockwise for updating faster reading, or turn it

clockwise for more resolution (digital display).

Pulling out the GATE TIME knob to freeze the latest displayed reading

for being written down. Then return the counter to normal operation by

pressing back the knob.

When the instrument is starting measurement, the LED indicator above

the GATE TIME knob will light. In normal operation, it will blink at a

rate setting by GATE TIME knob.

(3)

 

Trigger Level Adjustment

Adjust the trigger voltage of channel A input signal by pulling and

turning the TRIG LEVEL knob for triggering. The trigger voltage is

variable over 

±

2.5V×ATT setting.

If push the knob, it can be functioned as auto-triggering.(this knob is

only available for channel A.)

Содержание GFC-8131H

Страница 1: ...n 4 4 4 5 4 PANEL INTRODUCTION 6 5 APPLICATION 9 6 MAINTENANCE 6 1 Standard Method for Calibration 6 2 Cleaning 11 11 12 FREQUENCY COUNTER USER MANUAL 1 1 PRODUCT INTRODUCTION The intelligent counters are measuring frequency and period from the range of 0 01Hz to 120MHz on channel A and 50MHz to 1 3 2 7 GHz on channel B The products also provide special features of high resolution and sensitivity ...

Страница 2: ... 700MHz 15mVrms Max to 1GHz 25mVrms Max to 1GHz 25mVrms Max to 2GHz 40mVrms Max to 1 3GHz 50mVrms Max to 2 7GHz FREQUENCY COUNTER USER MANUAL 3 Coupling AC Impedance 50Ω Max Input Level 3Vrms sine wave General Resolution Gate time displays at least 7 digits for 1s 6 digits for 100ms 5 digits for 10ms respectively The maximum resolution for frequency measurement is 100nHz for 1Hz and 0 1Hz for 100M...

Страница 3: ...cated in the label attached on the real panel to replace correct fuses WARNING To avoid electrical shock the power cord protective grounding conductor must be connected to ground When line voltages are changed replace the required fuses shown as below Line voltage Range Fuse Line voltage Range Fuse 100V 120V 90 110V 108 132V T200mA 250V 220V 230V 198 242V 207 253V T100mA 250V WARNING To avoid pers...

Страница 4: ...level by setting on the TRIG LEVEL 9 TRIG LEVEL KNOB Pull the knob the TRIGGER LEVEL control will be variable over 2 5V ATT and push the knob to proceed auto set function 10 LPF ON Insert a 100kHz low pass filter into input A 11 ATT 1 20 Select attenuation for input A 1 Directly connect input signal to input amplifiers 20 Attenuate input signal by a factor of 20 12 COUP DC AC Select DC or AC coupl...

Страница 5: ... TIME adjustment affects the sampling rate and the resolution of the reading Turn the knob counterclockwise for updating faster reading or turn it clockwise for more resolution digital display Pulling out the GATE TIME knob to freeze the latest displayed reading for being written down Then return the counter to normal operation by pressing back the knob When the instrument is starting measurement ...

Страница 6: ...6 MAINTENENCE The following instructions are executed by qualified personnel only To avoid electrical shock do not perform any servicing other than the operating instructions unless you are qualified to do so 6 1 Standard method for calibration 1 Input offset voltage adjustment Push the TRIG LEVEL knob and turn it to the central position Connect a 10MHz sine wave to channel A input BNC Adjust SVR1...

Страница 7: ...in a solution of mild detergent and water Do not spray cleaner directly onto the instrument because it may leak into the cabinet and cause damage Do not use chemicals containing benzine benzene toluene xylene acetone or similar solvents Do not use abrasive cleaners on any portion of the instrument ...

Отзывы: