  FREQUENCY COUNTER

USER MANUAL

6

4. PANEL INTRODUCTION

(1). Power ON/OFF

Power on or off by using the button.

(2). Reset

Restart counting by resetting counter to zero.

(3). FREQ A

Select frequency mode of operation for input A.

(4). PRID A

Select period mode of operation for input A.

(5)

FREQ B

Select frequency mode of operation for input B.

(6). Gate Time(LED)

When Gate time LED is on, the counter’s main gate
circuit is open and the measurement is in processing.

(7). Gate Time(KNOB) Select variable measurement time from 10ms to 10s

continuously (minimum=1 period of the input signal).
The display value can be hold by pulling the knob until
it is pushed back.

(8). TRIG LEVEL

(LED)

Indicate the input signal above or below the trigger level
by setting on the TRIG LEVEL.

(9). TRIG LEVEL

(KNOB)

Pull the knob, the TRIGGER LEVEL control will be
variable over 

±

2.5V × ATT and push the knob to

proceed auto-set function.

(10)
.

LPF/ON

Insert a 100kHz low pass filter into input A.

(11) ATT ×1/×20

Select attenuation for input A

×1      Directly connect input signal to input amplifiers.

×20    Attenuate input signal by a factor of 20.

(12) COUP DC/AC

Select DC or AC coupling for input A.

  FREQUENCY COUNTER

USER MANUAL

7

(13) INPUT A

Input BNC for input A

(14) INPUT B

Input BNC for input B.

(15) S

The unit of displayed data is second (S).

(16) Hz

The unit of displayed data is Hertz (Hz).

(17) Exponent (LED)

Indicate the value of the measurement exponent as
shown below:
k=1000     M=1,000,000      G=1,000,000,000
m=1/1000   

μ

=1/1,000,000      n=1/1,000,000,000

(18) DISPLAY(LED)

8 digits red LED display.

(19) OVFL(LED)

OVFL (overflow) indicator shows that one or more of
the most significant digits are not displayed.

Содержание GFC-8131H

Страница 1: ...n 4 4 4 5 4 PANEL INTRODUCTION 6 5 APPLICATION 9 6 MAINTENANCE 6 1 Standard Method for Calibration 6 2 Cleaning 11 11 12 FREQUENCY COUNTER USER MANUAL 1 1 PRODUCT INTRODUCTION The intelligent counters are measuring frequency and period from the range of 0 01Hz to 120MHz on channel A and 50MHz to 1 3 2 7 GHz on channel B The products also provide special features of high resolution and sensitivity ...

Страница 2: ... 700MHz 15mVrms Max to 1GHz 25mVrms Max to 1GHz 25mVrms Max to 2GHz 40mVrms Max to 1 3GHz 50mVrms Max to 2 7GHz FREQUENCY COUNTER USER MANUAL 3 Coupling AC Impedance 50Ω Max Input Level 3Vrms sine wave General Resolution Gate time displays at least 7 digits for 1s 6 digits for 100ms 5 digits for 10ms respectively The maximum resolution for frequency measurement is 100nHz for 1Hz and 0 1Hz for 100M...

Страница 3: ...cated in the label attached on the real panel to replace correct fuses WARNING To avoid electrical shock the power cord protective grounding conductor must be connected to ground When line voltages are changed replace the required fuses shown as below Line voltage Range Fuse Line voltage Range Fuse 100V 120V 90 110V 108 132V T200mA 250V 220V 230V 198 242V 207 253V T100mA 250V WARNING To avoid pers...

Страница 4: ...level by setting on the TRIG LEVEL 9 TRIG LEVEL KNOB Pull the knob the TRIGGER LEVEL control will be variable over 2 5V ATT and push the knob to proceed auto set function 10 LPF ON Insert a 100kHz low pass filter into input A 11 ATT 1 20 Select attenuation for input A 1 Directly connect input signal to input amplifiers 20 Attenuate input signal by a factor of 20 12 COUP DC AC Select DC or AC coupl...

Страница 5: ... TIME adjustment affects the sampling rate and the resolution of the reading Turn the knob counterclockwise for updating faster reading or turn it clockwise for more resolution digital display Pulling out the GATE TIME knob to freeze the latest displayed reading for being written down Then return the counter to normal operation by pressing back the knob When the instrument is starting measurement ...

Страница 6: ...6 MAINTENENCE The following instructions are executed by qualified personnel only To avoid electrical shock do not perform any servicing other than the operating instructions unless you are qualified to do so 6 1 Standard method for calibration 1 Input offset voltage adjustment Push the TRIG LEVEL knob and turn it to the central position Connect a 10MHz sine wave to channel A input BNC Adjust SVR1...

Страница 7: ...in a solution of mild detergent and water Do not spray cleaner directly onto the instrument because it may leak into the cabinet and cause damage Do not use chemicals containing benzine benzene toluene xylene acetone or similar solvents Do not use abrasive cleaners on any portion of the instrument ...

Отзывы: