REMOTE CONTROL
183
[SENSe:]AVERage:TCONtrol
Selects the digital filter.
Parameter: MOV | REP
Example: SENS:AVER:TCON MOV
Sets the digital filter to the Moving filter.
[SENSe:]AVERage:TCONtrol?
Returns the current digital filter type.
Return parameter: MOV (moving), REP (repeating)
[SENSe:]AVERage:COUNt
Sets the digital filter count.
Parameter: <NR1> (2 ~ 100) | MIN | MAX
Example: SENS:AVER:COUN 100
Sets the digital filter count number to 100.
[SENSe:]AVERage:COUNt?
Returns the digital filter count.
Return parameter: <NR1> (+002~+100)
[SENSe:]AVERage:WINDow
Selects a digital filter window
Parameters: 0.01 | 0.1 | 1 | 10 | NONE
Example: SENS: AVER: WIND 0.1
Sets the digital filter window to 0.1%
[SENSe:]AVERage:WINDow?
Return the current digital filter window value
Return parameters: 0.01, 0.1, 1, 10, NONE
[SENSe:]AVERage:STATe
Turns the digital filter On/Off.
Parameter: ON | OFF
Example: SENS:AVER:STAT ON
Turns the digital filter on.
[SENSe:]AVERage:STATe?
Returns the state of the digital filter (on or off).
Return parameter: 0|1, 0=OFF, 1=ON