REMOTE CONTROL
181
[SENSe:]TEMPerature:RJUNction:SIMulated
Set temperature simulation value.
Parameter: <NRf>(0.00 ~ 50.00)
Example: SENS:TEMP:RJUN:SIM 25.00
Sets the thermocouple junction temperature to 25
˚
C.
[SENSe:]TEMPerature:RJUNction:SIMulated?
Returns temperature simulation value.
Return parameter: <NR1> (+0000~+5000) ,where +0000=0.00
˚
C,
+5000=50.00
˚
C
[SENSe:]TEMPerature:RTD:TYPE
Sets the 2W RTD sensor type.
Return parameter: Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)
Example: SENS:TEMP:RTD:TYPE PT100
Sets the 2W RTD sensor to PT100
[SENSe:]TEMPerature:RTD:TYPE?
Returns the 2W RTD sensor type.
Return parameter: PT100, D100, F100, PT385, PT3916, USER
[SENSe:]TEMPerature:RTD:ALPHa
Sets the 2W RTD Alpha coefficient.
Parameter: <NRf> (0~10)
Example: SENS:TEMP:RTD:ALPH 0.00385
[SENSe:]TEMPerature:RTD:ALPHa?
Returns the 2W RTD Alpha coefficient.
[SENSe:]TEMPerature:RTD:BETA
Sets the 2W RTD BETA coefficient.
Parameter: <NRf> (0~10)
Example: SENS:TEMP:RTD:BETA 0.00495
[SENSe:]TEMPerature:RTD:BETA?
Returns the 2W RTD BETA coefficient.
[SENSe:]TEMPerature:RTD:DELTa
Sets the 2W RTD DELTa coefficient.
Parameter: <NRf> (0~10)
Example: SENS:TEMP:RTD:DELT 0.0000568
[SENSe:]TEMPerature:RTD:DELTa?
Returns the 2W RTD DELTa coefficient.