Operator's guide
UNLOCK
This command unlocks the sensor mass and prepares the instrument to
begin operating.
If
UNLOCK
is activated when the mass is already unlocked, the
processor will lock it and aiempt to unlock again. This is useful if you
suspect that the locking procedure has failed.
Again, you can use the
BUSY
LED to monitor the progress of unlocking.
1. The instrument checks to see whether the sensor mass is locked.
2. The instrument checks the sensor base.
3. The mass is unlocked.
4. The sensor base is unlocked. This process takes longer.
Afer u nlocking, t he i nstrument a utomatically p erforms a r ound o f
centring (see below).
CENTRE
This command re-centres the mass. If the mass is clamped, or if the
sensor mass position does not exceed ± 1.2 V, the
CENTRE
command
does nothing. Otherwise, it aiempts to zero the output of the sensor by
exerting small extra forces on the boom. A motor-driven adjuster presses
a small spring lever against the boom until the mass position sensor
indicates an ofset close to zero.
December 2004
53