For boreholes with a metallic casing at the boiom and plastic above, we
recommend connecting a cable between the sensor housing and the
ground plate so that the lower borehole casing acts as the earthing point.
Again, the
If there is a signifcant distance ((30 m) between the borehole and the
surface installation, the resistance of the earth cable may make it
impractical to use the borehole as an earthing point. In these cases, you
will have to connect the local ground plate to an earth stake near to the
enclosure; any coupling between this sensor-local earth line and ground
lines for other parts of the system must be minimized.
Installations with DC power supplies
Güralp sensors require a 24 V DC power supply. In most cases, this is
provided by an isolating DC/DC converter installed at the surface. This
converter can be earthed to the local ground plate as above.
However, DC/DC converters contain sensitive electronics, which must
be protected thoroughly. We recommend installing a full surge
protection unit in addition to the spark gap protector. This protection is
installed on the supply side of the isolator, so it must be earthed
separately
from the borehole installation. Otherwise, transients in the
42