11
Auto apical point reverse
The file will reverse when the file tip reaches the apical point.
Auto apical point stop
The file will stop when the file tip reaches the apical point.
Disable apical action function
The file rotating as usual even if reach the reference apical point.
2.8.1 The device contains a file library with the preset popular NiTi file
systems.
2.8.2 Please follow the file manufacturer’s instructions for use .The file
system shown on the display must always match the file in use.
2.8.3 Torque and speed values are subject to change by the file
manufacturers without notice. Therefore, the preset values in the library
must be checked prior to use.
2.8.4 Please use the 8 sets Individual Program to create your own file
sequence. This enables you to manage your own series of files.
2.8.5 Never use Safety Glide Path files in continuous rotary mode. Never
use rotating files in Safety Glide Path mode.
2.8.6 Safety Glide Path mode
Adjustable every 10 degrees, adjustment range: 20º-400º.
The rotation angle is adjustable, but the forward angle must be equal
to the reverse angle.