33
Guidance & Declaration - electromagnetic immunity
The model is intended for use in the electromagnetic environment specified below. The
customer or the user of the model should assure that It is used in such an environment
.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic environment
- guidance
E l e c t r o s t a t i c
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
tansient/burst IEC
61000-4-4
±2kV for power
supply lines
±1 kV for lnput/
output lines
±2kV for power
supply lines
±1kV for
interconnecting
cable
Mains power quality should be
that of a typical commercial or
hospital environment.