![Guildline 6622T Series Скачать руководство пользователя страница 84](http://html1.mh-extra.com/html/guildline/6622t-series/6622t-series_operation-manual_2278477084.webp)
Section 5
OM6622T-A-00
30 January
2018
5-4
5.2.
TEMPERATURE RATIO VERIFICATION
This section outlines the recommended thermometry based ratio verification requirements for the
model 6622T.
5.2.1.
Purpose
The purpose of the procedure in this section is to assure that the model 6622T Series Bridge
functions properly and meets the manufacturer’s additional specifications for Temperature
Measurement. It is recommended that these procedures be performed after any corrective
maintenance activity, or if the operator perceives a problem.
It is recommended that all of these procedures be performed on an annual basis to verify
performance of the model 6622T for its recertification.
5.2.1.
Equipment Required
The following equipment is required for this procedure.
-
10 ohm standard resistor, stable to ±0.01 ppm over a 24 hour period (x2)
-
100 ohm standard resistor, stable to ±0.01 ppm over a 24 hour period (x2)
-
1k ohm standard resistor, stable to ±0.01 ppm over a 24 hour period (x2)
-
25 ohm standard resistor, stable to ±0.01 ppm over a 24 hour period
-
300 ohm standard resistor, stable to ±0.2 ppm over a 24 hour period
-
Temperature controlled environment sufficient to maintain the reference and working standard
resistors to their specified uncertainties
Note: Ten of these standard resistors are available in the optional model 6634A
Resistance Standard and up to four reference resistors are available in the
portable Guildline Instruments model 6634TS Resistance Standard.
Содержание 6622T Series
Страница 2: ......
Страница 8: ...OM6622T A 00 30 January 2018 vii...
Страница 10: ...OM6622T A 00 30 January 2018 ix...
Страница 16: ...Section 1 OM6622T A 00 30 January 2018 1 6...
Страница 34: ...Section 3 OM6622T A 00 30 January 2018 3 8 Figure 3 3 Rear Panel Connections Low Ohm Mode with 6623...
Страница 35: ...Section 3 OM6622T A 00 30 January 2018 3 9 Figure 3 4 Rear Panel Connections Low Ohm Mode with 6623A 300...
Страница 37: ...Section 3 OM6622T A 00 30 January 2018 3 11 Figure 3 5 Rear Panel Connections PRT SPRT...
Страница 80: ...Section 4 OM6622T A 00 30 January 2018 4 42...
Страница 92: ...Section 6 OM6622T A 00 30 January 2018 6 6...
Страница 114: ...Section 7 OM6622T A 00 30 January 2018 7 22...