User Manual
33
100 to 800°C
•
If the target temperature is beyond the current filter range, the symbol “+++” or “---” will
appear in the temperature display bar under the live image. It indicates that the filter range
should be changed manually
•
To change the filter range manually, go to the live thermal image, and press button A for
an instant to select the temperature setup menu at the right side of the image. Move the
joystick up or down to select option filter and then right or down to change its value.
Note:
In auto modes, when the target temperature is higher than 250°C and the Center
cursor aims at the target, the camera will automatically switch to filter range 2. In manual
mode, filter range shall be changed manually.
•
After changing the option values, press button C to save the change and exit the menu
•
Using the touch pen can also do the above operation. Click the temperature setup menu
with the pen first, move the joystick left or right to change option values, and then click the
menu again with the pen. Now you have saved the changes and returned to the live
image.
Basic Temperature Measurement
For basic temperature measurement, first ensure that a suitable temperature range is set and
that the image is in focus. Two methods are then available,
dynamic real-time temperature
measurement
and
static temperature measurement
Dynamic real-time temperature measurement
•
This method is always available in live thermal image.
•
There are two cursors on the live thermal image. One is always fixed in the centre of
image; the other is always tracing the hottest spot of the image.
Содержание ThermoPro TP8
Страница 15: ...User Manual 15 ...
Страница 17: ...User Manual 17 1 LCD mount 1 Tightening gear 2 Touch pen 3 Contact pins at the LCD screen foot ...
Страница 49: ...User Manual 49 The submenus can be navigated by either clicking with the touch pen or moving the ...
Страница 79: ...User Manual 79 proceed further Installation continues and finishes quickly Click button Finish ...