16
If the dynamic time (
t
dyn
) exceeded the normative 0.75 seconds, the LED
within the second display segment flashes.
If the force value at the end of the 5s measurement period (
F
End
) exceeded 25
N, the LED within the third display segment flashes.
Measurement result:
F
dyn
= 326 N
t
dyn
> 0,75 s
Measurement result:
F
dyn
= 326 N
F
End
> 25 N