13
ADICOS HTL-2
410-2410-008 EN20
MAINTENANCE
8 Malfunctions
Description of malfunction
Possible cause
Solution
Operating LED and heating
LED do not light up
Batteries are empty Charge the SOLO battery baton
(–› Chapter 5.2)
Operating LED and heating
LED do not light up
Battery baton
improperly inserted
Remove the SOLO Akkustange
and connect it again
(–› Chapter 5.3)
Operating LED and
heating LED blink
Low battery power
Charge the SOLO battery baton
(–› Chapter 5.2)
Heating LED does not go out
Low battery power
Charge the SOLO battery baton
(–› Chapter 5.2)
9 Maintenance
9.1 Cleaning
Cleaning the HTL test tube
1
Blow out the test tube with compressed air if it gets dirty
2
Clean the hot plate with a moist towel
Do not use any aggressive cleaning agents!
Cleaning the HTL case
►
Clean the outer surfaces of the HTL-2 with a moist towel as needed
Do not use any aggressive cleaning agents!
Maintaining the HTL swivel joints
1
If it becomes difficult to rotate the HTL-2 swivel joints, blow them out with compressed air
2
Grease the swivel joints with penetrating oil (e.g. WD-40)
9.2 Storage
Store the device in its carrying case when it is not in use.