Chapter 4
This chapter includes:
When to calibrate the TympStar
page 4-2
Calibration requirements
page 4-2
Preparing for calibration
page 4-4
Cautions and warnings
page 4-4
Tools and equipment required
page 4-4
Cleaning the probe
page 4-4
Cal/Normal switch
page 4-6
Cal Options switch
page 4-6
Probe serial number
page 4-26
Resolving problems
page 4-27
Calibration procedures
page 4-28
TympStar calibration controls
page 4-28
Custom transducers
page 4-32
Transducer connections
page 4-32
Version 1 calibration
page 4-33
Version 2 calibration
page 4-51
Returning to Normal Test mode
page 4-66
The TympStar must be calibrated when the digital, PC104 or analog board is
repaired, if a transducer is changed or if the probe is repaired. Probe tone fre-
quency data (Self-Cal data) are stored in the TympStar instrument. All other
data for transducers and pressure are stored in the probe. Normally, circuit
boards or the probe can be replaced with new complements from the factory
without the need for calibration.
GSI recommends quarterly calibration checks for the GSI TympStar along with
annual certification. ASHA requires quarterly electro-acoustic calibration
checks and annual electro-acoustic calibration. It is good practice to perform
daily biologic checks.
Some aspects of TympStar calibration can be checked singularly without the
need for a complete or routine calibration sequence by simply navigating
through the calibration menus to the desired screen and performing the check.
However, calibration is typically as a complete calibration sequence
or as a routine calibration sequence.
A complete calibration sequence must be performed if the following occurs:
• A performance problem is suspected
• A repair has been made to the TympStar involving the digital board,
analog board, PC104 board, probe box or transducers
The complete calibration sequence must include the following:
• Loading default data and performing Self-Cal
• Contra, Ipsi and probe tone SPL calibrations
• Y (Version 1) or Y/B/G (Version 2) calibration
• Pressure system calibration
Grason-Stadler
When to calibrate
the TympStar
Calibration require-
ments
Complete calibration
sequence
4 - 2
Содержание TYMPSTAR 1
Страница 1: ...Part Number 2000 0110RevB...
Страница 20: ...Chapter 1 1 8 Grason Stadler...
Страница 72: ...Chapter 3 Grason Stadler 3 32...
Страница 168: ...Chapter 5 Grason Stadler 5 30...
Страница 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Страница 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Страница 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Страница 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Страница 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Страница 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Страница 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Страница 178: ...Chapter 6 Grason Stadler 6 10...
Страница 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Страница 180: ...Chapter 6 Grason Stadler 6 12...
Страница 185: ...Hardware Description Interconnection diagram GSI TympStar Version 1 and Version 2 Service Manual 7 5...
Страница 190: ...Chapter 7 Printer board Digital board Power supply board Pump Analog board PC 104 board Grason Stadler 7 10...