Troubleshooting
Error Messages
Error messages are text and are generated by the Control Processor. The following
messages may occur either as the result of an improper selection made by the op-
erator or by a fault internal to the GSI TympStar:
ERROR INVALID SELECTION
An unlabeled softkey was pressed.
ERROR INVALID SELECTIONWHILE CALIBRATINGCOMPLIANCE
The altitude up/down keys were pressed, or the Y calibration screen attempted
printing while calibrating the compliance.
INVALID INTENSITY LIMIT REACHED
A stimulus level (dB HL) was selected that was either outside the legally set limits or
would require an unattainable attenuator setting to produce the stimulus level.
NOTE PROBLEM STORING DATA- CONTACT SERVICE
If this message is reoccurring, the digital board is most likely at fault. A Control
Processor HELP 1052 message is also likely to occur.
ERROR INVALID SELECTION DURING CAL MODE
The key pressed is not valid while in the calibration mode.
NOTE SITE ALTITUDE DEFAULTED TO 0 - RE-ENTER IFNECESSARY
If this message occurs, the altitude cal procedure should be attempted.
ERROR INVALID SELECTION UNTILA CAL MODE SELECTED
The key pressed is not valid in the current calibration mode, which is indicated by
the current calibration screen displayed.
ERROR INVALID SELECTION DURING IPSI/CONTRA SPL CAL
The key pressed is not valid while in IPSI/CONTRA SPL calibration mode.
ERROR INVALID SELECTION DURING PROBE SPL CAL
The key pressed is not valid while in the probe SPL calibration mode.
ERROR INVALID SELECTION DURING COMPLIANCE CAL
The key pressed is not valid while in the compliance calibration mode.
ERROR INVALID SELECTION DURING PRESSURE CAL
The key pressed is not valid while in the pressure calibration mode.
GSI TympStar Version 1 and Version 2 Service Manual
8 - 19
Содержание TYMPSTAR 1
Страница 1: ...Part Number 2000 0110RevB...
Страница 20: ...Chapter 1 1 8 Grason Stadler...
Страница 72: ...Chapter 3 Grason Stadler 3 32...
Страница 168: ...Chapter 5 Grason Stadler 5 30...
Страница 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Страница 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Страница 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Страница 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Страница 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Страница 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Страница 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Страница 178: ...Chapter 6 Grason Stadler 6 10...
Страница 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Страница 180: ...Chapter 6 Grason Stadler 6 12...
Страница 185: ...Hardware Description Interconnection diagram GSI TympStar Version 1 and Version 2 Service Manual 7 5...
Страница 190: ...Chapter 7 Printer board Digital board Power supply board Pump Analog board PC 104 board Grason Stadler 7 10...