Should a load cell become damaged due to an overload condition, replacement load cells can be
ordered from GSE Scale Systems according to the part numbers in
Table 1
on page 3.
1. Unbolt the defective load cell, and then remove leveling load foot from the load cell.
2. Disconnect load cell cable from junction box and attach a chase wire to the load cell cable.
DO NOT CUT THE LOAD CELL CABLE!
3. Pull the load cell cable out of the formed frame channels. The chase wire will be used to pull
the new load cell cable through the formed channel.
4.
Install the new load cell by reversing the preceding steps.
5. To verify correct load cell/junction box terminal matching, see the numbers on the terminals
inside the junction box and the corner numbering diagram in
Error! Reference source not
found.
.
6. Position load cell and torque down with a wrench; tighten all mounting bolts to 75 ft./lbs.
7. Route the load cell cables in each corner so that the cable is free from possible contact with
the load cell or load foot. Pull cable tight towards junction box, coil up extra length with cable
ties and securely store inside channel.
8. Pass end of load cell cable through cable fittings in the NEMA 4X junction box. Corner
correction trimming and calibration is necessary after load cell replacement. Follow the
calibration instructions on page 1.
Table 1: Load Cell Part Numbers
Size Load
Cell
Capacity Part Number
Platform
19 x 25
24 x 30
500 lb
E0-50459
PT 800 / PT 820 CS
1K E0-50460
2K E0-50460
19 x 25
24 x 30
500 lb
E0-50459
PT810 / PT 830 SS
1K E0-50460
2K E0-50462