English (GB)

4

5.2 Saving data, important notes

The batteries save data (stored results and photometer setting).
During battery change, the data in the photometer is saved for 
two minutes.
If the change time exceeds two minutes, all stored data and 
settings will be lost, and the procedure described in 5.1 must be 
executed once.

5.3 Replacement of batteries

Fig. 2

Replacement of batteries (back)

1. Switch off the photometer.
2. Remove the vial from the sample chamber.
3. Place the photometer upside down on a clean and even 

surface.

4. Unscrew the four screws (A) of the battery compartment 

cover (B).

5. Lift off the battery compartment cover (B) at the notch (C).
6. Remove old batteries (D).
7. Position four new batteries.

– Ensure the correct polarity!

8. Check the seal ring (E) at the battery compartment cover (B) 

and make sure that it is tight-fitting.

9. Place the battery compartment cover (B) back on the 

photometer without moving the seal ring (E).

10. Tighten the screws carefully.

6. Operation

The multilingual plain-text operator prompting of the DIT-M 
photometer offers high operating convenience. The user can 
choose the following languages: German, English, French, 
Spanish, Italian, Portuguese, Polish.
Two operating modes can be selected:
• In normal mode all steps of the measurement are displayed.
• The expert mode for the proficient user shows shortened texts 

and does not request every single step, which saves time.

See also section 

6.3.5 Lab function

.

6.1 Function keys

6.1.1 Overview of function keys

6.1.2 Displaying time and date

Note

For replacement, have a screwdriver and new 
batteries at hand.

Caution

Although the screws (A) are dismantled, the battery 
compartment cover (B) provides resistance against 
opening due to the special sealing.
For opening don't use tools, which can cause 
damage to the device!

T

M

04

 16

64

 27

11

Caution

To ensure that the photometer is waterproof, the seal 
ring (E) must be in position and the battery 
compartment cover (B) must be fixed with the four 
screws (A).

Note

EC guideline 2006/66/EC requires users to return all 
used and worn-out batteries and accumulators. They 
must not be disposed of together with ordinary 
domestic waste.

Note

Do not use rechargeable batteries (accumulators)!

AA/LR6

AA/LR6

A

A/

LR

6

AA/L

R6

(E) Seal ring

(B) Battery 

compartment 

cover

(A) Screws

(C) Notch

(D) Batteries

(F) Back

Press the ["clock"] key.

The display shows time and date.

After 15 seconds, the photometer returns 
automatically to the previous display. You 
can also press the 

 or [Esc] keys.

 

 

 

 

 

 

 

 

 

 

 

 

4

7

9

6

.

0

3

2

5

8

1

On

Off

Shift

Esc

F 1

F 2

F 3

Mode

Store

Zero

Test

Switch the photometer on or off.

Obtain the number keys 0-9 by pressing the shift key. 

Hold down the shift key and press the required number 

key(s). Example: [Shift] + [1][1].

Return to selection of methods or to the previous menu.

Function key: description in the text where the key is 

used.

Function key: description in the text where the key is 

used.

Function key: description in the text where the key is 

used.

Confirm.

Menu of photometer settings and other functions.

Store a displayed test result.

Perform zero setting.

Perform a test.

Display date and time / user countdown.

Decimal point.

Move the cursor up or down.

19:27:20 2009-06-15

Esc

Содержание DIT-M

Страница 1: ...DIT M Photometer Installation and operating instructions GRUNDFOS INSTRUCTIONS Other languages http net grundfos com qr i 95727428 Material safety data sheets http net grundfos com qr i 98351242...

Страница 2: ...0 3 mg l Al 20 7 5 Bromine 0 05 13 mg l Br2 21 7 6 Chloride 0 5 25 mg l Cl 21 7 7 Chlorine 0 01 6 mg l Cl2 22 7 8 Chlorine dioxide 0 02 11 mg l ClO2 23 7 9 Chlorine dioxide analysis ClO2 chlorite Cl2...

Страница 3: ...keys Shift 1 and press key to start initialization of storage system Warning Do not open or dismantle the photometer Cleaning maintenance and repair must only be carried out by authorised personnel TM...

Страница 4: ...Displaying time and date Note For replacement have a screwdriver and new batteries at hand Caution Although the screws A are dismantled the battery compartment cover B provides resistance against ope...

Страница 5: ...0 2 0 0 Confirm with the key The set value is displayed To start countdown press the key After countdown is finished the photometer returns automatically to the previous display Press the Shift F1 ke...

Страница 6: ...wn mode description The vial must not be placed in the sample chamber Press the Test key Prepare the water sample as described in the respective method description and place the vial in the sample cha...

Страница 7: ...isplay appears Press the Esc key to return to the method selection It is also possible to enter the required method number directly such as Shift 1 6 0 for CyA TEST cyanuric acid Confirm your input wi...

Страница 8: ...od 33 Displaying only results of one selected method 12 Print 20 Printing all stored results 10 Print Code No 22 Printing only results of a selected Code No range 10 Print date 21 Printing only result...

Страница 9: ...p is switched off Press the Mode key Press the Shift and 1 0 keys Confirm with the key This display appears To select the required language from the displayed list press the arrow key or Confirm with...

Страница 10: ...May 19 2009 Shift 0 9 0 5 1 9 Confirm with the key This display appears To print all stored results in the selected date range press the key After printing the photometer returns automatically to the...

Страница 11: ...Baud rate is 19200 select save cancel Esc Esc Shift 0 Mode 3 Storage display all data Start cancel Esc print F3 print all F2 no data Press the Mode key Press the Shift and 3 1 keys Confirm with the ke...

Страница 12: ...t to delete the data cancel and exit the menu by pressing the Esc key Shift Storage 3 Mode 3 Storage Method 80 Bromine Start cancel Esc print F3 print all F2 20 Acid demand 40 Aluminium T 80 Bromine T...

Страница 13: ...Aluminium T 0 1 0 2 mg l Al 80 Bromine T calibration via basic method 100 Chlorine free 100 Chlorine T 0 5 1 5 mg l Cl2 120 Chlorine dioxide T calibration via basic method 100 Chlorine free 90 Chlori...

Страница 14: ...ns 1 10 are displayed in the method list even if they are not programmed by the user Non programmed user methods cannot be activated User method list switch on all methods This mode function activates...

Страница 15: ...To start the entry mode press the Shift and 1 keys Enter the required wavelength by pressing the appropriate number key such as Shift and 2 for 560 nm Shift Mlist all off switch off all methods YES 1...

Страница 16: ...S2 accepted S3 ________ ESC F1 Store stored Shift User polynoms choose No ______ 800 824 5 Mode 6 0 8 0 Overwrite polynom YES 1 NO 0 wavelength 1 530 nm 2 560 nm 3 610 nm 4 430 nm 5 580 nm 6 660 nm 2...

Страница 17: ...Press the Mode key Press the Shift and 8 0 keys Confirm with the key This display appears To increase the contrast of the LC display press arrow key To decrease the contrast of the LC display press ar...

Страница 18: ...entering the sample chamber Correct position of the vial 24 Fig 4 Correct position of the vial 24 6 Bubbles on the inside wall of the vial result in incorrect measurements To prevent this remove the b...

Страница 19: ...727747 7 8 4 100 Chlorine dioxide analysis T Chlorine dioxide chlorite free chlorine combined chlorine 0 01 6 mg l in mg l 0 1 0 04 1 2 0 08 2 3 0 18 3 4 0 25 4 6 0 35 0 01 250 DPD No 1 tablets 957277...

Страница 20: ...ised water 2 To get accurate results the sample temperature must be between 20 C and 25 C 3 A low test result may be given in the presence of fluorides and polyphosphates The effect of this is general...

Страница 21: ...ave different effects on the precipitation reaction 3 Ions which also form deposits with silver nitrate in acidic media such as bromides iodides and thiocyanates interfere with the measurement 4 If ne...

Страница 22: ...the same way as chlorine 7 7 3 Chlorine differentiated 7 7 4 Chlorine free The following selection is shown in the display for the differentiated determination of free combined and total chlorine for...

Страница 23: ...lorine dioxide as chlorine reading from the quoted total chlorine reading 3 See also section 7 8 Chlorine dioxide 0 02 11 mg l ClO2 Measurement 1 Fill a clean vial 24 with 10 ml of the water sample cl...

Страница 24: ...and swirl gently several times until the tablet is dissolved 22 Place the vial in the sample chamber correctly See fig 4 23 Press the Test key Wait for a reaction period of 2 minutes When the reaction...

Страница 25: ...o press ZERO 24 Chlorine diff free total Zero accepted prepare Test press TEST 20 Place the vial correctly in the sample chamber See fig 4 21 Press the Test key Write down the result as reading A 22 P...

Страница 26: ...the sample the escape of hydrogen peroxide gases e g by pipetting or shaking must be avoided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried o...

Страница 27: ...y Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l hydrogen peroxide TM04 1660 3510 prepare...

Страница 28: ...determination of ozone in the presence of with chlorine for the determination of ozone in absence of without chlorine Select the desired differentiation with without chlorine using the arrow keys and...

Страница 29: ...PD No 1 tablet No 95727747 and one DPD No 3 tablet No 95727750 straight from the foil and crush the tablets using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly wit...

Страница 30: ...es with salt contents as follows The values of Parson and Douglas 1926 are based on the use of Clark and Lubs buffers 1 mol NaCl 58 4 g l 5 8 Indicator Salt content of the sample Phenol red 1 molar 2...

Страница 31: ...Zero is incorrect Make sure that the zero vial has been done Insert zero vial repeat the zero setting Clean sample chamber repeat zero The calculation of a value such as combined chlorine is not possi...

Страница 32: ...727773 Abbreviation Definition C degree Celsius Centigrade F degree Fahrenheit F C x 1 8 32 dH degree German Hardness fH degree French Hardness eH degree English Hardness aH degree American Hardness A...

Страница 33: ...one di Conformit CE Grundfos dichiara sotto la sua esclusiva responsabilit che i prodotti DIT L DIT M e DIT IR ai quali si riferisce questa dichiarazione sono conformi alle seguenti direttive del Cons...

Страница 34: ...34...

Страница 35: ...ia S r l Via Gran Sasso 4 I 20060 Truccazzano Milano Tel 39 02 95838112 Telefax 39 02 95309290 95838461 Japan GRUNDFOS Pumps K K 1 2 3 Shin Miyakoda Kita ku Hamamatsu 431 2103 Japan Phone 81 53 428 47...

Страница 36: ...splayed in this material including but not limited to Grundfos the Grundfos logo and be think innovate are registered trademarks owned by The Grundfos Group All rights reserved 2019 Grundfos Holding A...

Отзывы: