6.3.5 Lab function
Reduced operator guidance in Profi-Mode
The following information is always stored in the methods:
1. method
2. range
3. date and time
4. differentiation of results
5. detailed operator guidance
6. compliance with reaction periods (countdown).
If the Profi-Mode is active, the photometer provides only a
minimum of operator guidance. The above items 4, 5, 6 are not
included.
Note:
Storage of test results is possible in the Profi-Mode. When results
are stored, the display shows "Profi-Mode" additionally.
The selected settings are kept by the photometer even when it is
switched off. To change the photometer setting, a new setting is
required.
6.3.6 User operations
User method list
After switching on the photometer, a scroll list of all available
methods is automatically shown in the display. To shorten this list,
the user can define a user method list according to his own
requirements.
After an update, new methods are automatically added to the
user method list.
The program structure requires that the user method list must
have at least one active (switched-on) method. For this reason,
it is necessary to activate first all required methods and then to
switch off the automatically activated one, if this method is not
required.
User method list, adaption
Note:
If only a few methods are required, it is recommended to perform
Mode 62 "Mlist all off" first, followed by Mode 60 "Method list".
Only the methods that are to appear in the method selection list,
have to be marked with the dot [
●
].
All user Polynomials (1-25) and Concentrations (1-10) are
displayed in the method list, even if they are not programmed by
the user. Non-programmed user methods cannot be activated.
User method list, switch on all methods
This mode function activates all methods. After switching on the
photometer, a scroll list of all available methods is automatically
shown in the display.
Press the [Mode] key.
Press the [Shift] and [5][0] keys.
Confirm with the
key.
This display appears.
To switch the Profi-Mode off,
press the [Shift] and [0] keys.
To switch the Profi-Mode on,
press the [Shift] and [1] keys.
This display appears.
Confirm with the
key.
Shift
<Profi-Mode>
0
Mode
5
switched off
or
switched on
ON: 1
OFF: 0
0
1
Press the [Mode] key.
Press the [Shift] and [6][0] keys.
Confirm with the
key.
This display appears.
To start, press the
key.
The complete method list is displayed.
Methods with a dot [
●
] behind the method
number will be displayed in the method
selection list. Methods without a dot will not
be displayed.
To select the required method from the
displayed list, press the arrow keys or .
To switch between "active" [
●
] and
"inactive" [ ], press the [F2] key.
Select the next method, set "active" or
"inactive" until all methods have the desired
setting.
Confirm and store with the
key.
To cancel without storing, press the [Esc]
key.
Press the [Mode] key.
Press the [Shift] and [6][1] keys.
Confirm with the
key.
This display appears.
• To display all methods in the method
selection list, press the [Shift] and [1]
keys.
• To keep the valid method selection list,
press the [Shift] and [0] keys.
The photometer returns automatically to the
"Mode" menu.
Shift
<Method list>
selected:
●
toggle: F2
save:
cancel: Esc
0
Mode
6
<Method list>
>> 20
●
Acid demand
40
●
Aluminium
80
●
Bromine
...
>> 40
●
Aluminium
F2
>> 40
Aluminium
F2
>> 40
●
Aluminium
Shift
<Mlist all on>
switch on all methods
YES: 1, NO: 0
1
Mode
6
1
0
Содержание DIT-M