Reverse osmosis system
AVRO 125 TS/TL
Order no. 185 752 945-inter Edited by: nkes-mrie G:\BA-752945-INTER_185_AVRO_125.DOCX
36
2.3 Access to the
programming
levels – change
parameters
1. Simultaneously pressing the and buttons (> 1 s) causes
the code request C 000 to appear.
2. Set the required code with or button and confirm with
button.
3. Within the programming level, select the desired parameter with
the or button and open it for editing with the button
(value starts flashing).
4. Use the or button to change the parameter to the required
value.
5. Save the new parameter setting with the button (value stops
flashing) or reject the change by simultaneously pressing the
and buttons, and close the parameter again (value stops
flashing, previous setting remains saved).
6. Return to the basic display time by simultaneously pressing
and buttons.
7. If no button is pressed during longer than 5 minutes within a
parameter level, the display automatically jumps back to the basic
time display. Any opened parameters (flashing value) and closed
and the old set value is maintained.
2.4 Software version
Display
Parameter
P1.00
Software version of the RO-matic control unit
3 | Programming levels
3.1 Input logic code 113
Display
factory
setting
Parameter Setting
range
Comments
E-A: 1
Contact type level "a"
0 ... 1
0 = NO contact
1 = NC contact
E-b: 0
Contact type level "b"
0 ... 1
0 = NO contact
1 = NC contact
E-c: 0
Contact type level "c"
0 ... 1
0 = NO contact
1 = NC contact
EHP: 2
Type of contact
pressure switch
negative pressure HP
(high-pressure pump).
0 ... 3
0 = NO contact
1 = NC contact
2 = NO contact with auto
restart
1)
3 = NC contact with auto
restart
1)