LABO VEN S.IN S.7-18.wpd
9 of 23
07/18
3-6
THERMOMETER
3-6.1
On all standard models a dial thermometer is included. Insert it into its spring steel
holder so that it extends at least three inches (3") into oven work space. The
thermometer can be installed through any of the vent ports in the top of the oven.
3-7
ELECTRICAL
3-7.1
All electrical connections should be made in accordance with the appropriate local
and national codes. Refer to NFPA 70
A
National Electric Code
@
(see Appendix A).
3-7.2
Properly size the electrical supply using information provided on the oven
nameplate. Electric supply must include a safety shut off such as a circuit breaker
or disconnect switch between your power supply and the equipment.
3-7.3
W here a plug is provided, it must not be removed or changed. A properly sized
circuit and receptacle must be installed.
3-7.3.1
120 Volt Units - Plug into grounded (3 prong) 120 volt outlet. Do not
remove ground prong from plug. If no grounded outlet is available,
have one installed prior to using this equipment.
3-7.3.2
240 Volt Units - A plug suitable for use with your existing 240 volt
receptacle m ust be installed on the cord provided. Make sure that
your existing receptacle and the plug installed provide a secure
ground.
3-7.4
The oven must be adequately grounded. W here a plug is provided, do not rem ove
the ground prong. Grounding wire must be sized in accordance with local codes.
W here more strict codes do not exist, refer to the NFPA 70
A
National Electrical
Code
@
(see Appendix A).