Page 40
PSL 5.0 Pump Station Level Controller
Optional PZ34T_F
Maximum Range:
32 ft. (10 m)
Minimum Range (Deadband):
16" (406.4 mm)
Operating Frequency:
46
KHz
Beam Angle:
8º
Temperature Compensation:
Automatic,
continuous
Operating Temperature:
-40° to 150ºF (-40° to 65ºC)
Maximum Operating Pressure:
20 psi (1.35 Bar)
Sensor Face:
Teflon
Sensor Body:
PVC (not exposed to tank contents)
Cable Length:
25 ft. (7.6 m) continuous RG62AU coaxial Optional 50 ft. (15 m)
continuous
Maximum Cable Length:
500 ft (152 m) RG62AU coaxial (splice)
Hazardous Rating:
CSA rated Intrinsically Safe Class I, Groups C,D, Class II, Groups E,F,G
with optional Intrinsic Safety Barrier.
~ 4.
7
5"
(120 mm)
OVERALL
3
.5"
8
9
mm
2
5
ft (7
.6
m) RG
6
2AU
COAX
I
AL CABLE
(
5
0 ft 1
5
m OPT
I
ONAL)
3
/4“
NPT
I
SOLAT
I
ON
COUPL
I
NG
(SUPPL
I
ED)
3
/4”
NPT
1
"
(2
5.4
mm)
BOLT C
I
RCLE D
I
AMETER
TEFLON FACE
FLANGE D
I
AMETER