Possible Causes:
Corrective Action:
METER READING ERRATIC
Sensor mounted too close to valve, pump or
elbow
• Change sensor placement. Recommended 6-10
diameters from elbows, and 30 diameters from
pumps, controlling valves, orifice plates,
nozzles or open pipe discharge
NO FLOW INDICATION
Not enough suspended particles or gases in the
fluid
• Relocate sensor in more turbulent pipe section.
Mount sensor at 12 o'clock position on
horizontal pipe
Coupling compound washed out, or sensor loose
on pipe
• Remount sensor
• Use Dow Corning Silicone #4
Power interruption. No flow.
• Check fuse/breaker. Confirm flow
METER READING TOO HIGH
Calibration error
• Review
UNITS/MODE
menu and Pipe
ID
Vibration or noise on the pipeline
• Install in another location.
Pipe is not full
• Remount Sensor on vertical pipe
Nearby velocity increasing device (pump, valve,
orifice plate)
• Relocate sensor >30 pipe diameters from
velocity increasing device
Local electrical noise
• Ensure all Flowmeter wiring is in METAL
conduit and sensor cable shield is connected to
Ground stud
Variable Speed Drive interference
• Follow Drive manufacturers wiring and
Grounding instructions
• Relocate Flowmeter electronics, Sensor and
wiring away from VSD