5
Safety requirements
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and
special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any
other climatic conditions than those stated under "Specification".
2. If the device is transported from a cold to a warm environment condensation may result in a failure of the device.
In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new
start-up.
3. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be
marked accordingly to avoid re-starting.
Operator safety may be at risk if:
- there is visible damage to the device.
- the device is not working as specified.
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
4.
Warning:
Do not use this product as safety or emergency stop device or in any other application where failure of
the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.
5. Any changes or repair of the device is not allowed.
Please return device to manufacturer for repair or maintenance.
6
Specification
Measuring range:
humidity:
0,0 ... 100,0 %RH (rec. range of application: 10 ... 90 %RH)
temperature:
-40,0 ... +100,0 °C (for short-time up to 120 °C)
Accuracy:
(at nominal temperature)
humidity:
±2.5 % (at range 10 ... 90 %RH)
temperature:
±0.5 °C
Sensors:
humidity:
capacitive polymer-humidity sensor
temperature:
Pt1000
Electronics:
PC board with evaluation of measuring data and memory for sensor
data (measuring range, calibration etc.) integrated in handle.
Nominal temperature:
25°C
Working temperature:
sensor head and tube:
-40 ... +100 °C (for short-time up to 120 °C)
handle and electronic:
-25 ... +60 °C
Relative humidity:
0 ... +100 %RH.
Storage temperature
:
-25 ... +70 °C
Unit connection:
1.2 m PVC connection cable with 6-pin screened lockable Mini-DIN-
socket.
Dimensions:
Sensor tube:
Ø14 x 119 mm,
plastic handle:
Ø19 x 135 mm
Weight:
approx. 90 g
EMC:
The device corresponds to the essential protection ratings estab-
lished in the Regulations of the Council for the Approximation of Leg-
islation for the member countries regarding electromagnetic compati-
bility (2004/108/EG).
Additional fault: <1%