GREISINGER ResOx5695 Скачать руководство пользователя страница 1

H84.0.16.6AC-02

 

 

 

 

Application Instructions 
ResOx5695 

– Measuring oxygen in protective atmospheres 

For quick and comfortable measuring of residual oxygen in  food packaging in modified atmosphere packaging (MAP). 

Scope of Supply 

Connection scheme acc. to intended use 

Nr 

#  

 

 

The pressure connection has to be connected to port “IN” 
of the gas sampling pump. Correct connections are 
necessary for pressure compensation in stiff packages 

 

1  

Instrument GMH 5695 incl.  battery 

 

    


    

Sensor housing with pressure 
compensation connection incl. Sensor 
GOEL 

 

Gas sampling pump GS 150 incl. battery 

 

Tube set (3 tubes, T-piece) 

 

2  

GDZ Filter 

 

GOG-N insertion needles Ø 0,9 mm 

 

1  

GOG-B: 40 pieces sampling stickers 

 

1  

Case GKK 1420 

 

Operating and Short form manuals for 
GMH 5695, operating manual GS 150 

 

Application instructions ResOx 5695  

The measurement assembly is optimized for needles of Ø0.9mm. Using smaller needles may reduce  the necessary 
sampling volume. Attention: This may have an influence at the measuring precision 

  
The sharp needles are causing a risk of injury! Handle carefully! 
 

 

Sampling time:   

 ca. 15 s 

Sampling volume:  

Min. 20 ml (with filter GDZ 29) 

Choosing the sensor element 

Sensor element 

GOEL 381  

Standard at ResOx 5695-L

 

GOEL 370 (Rev 2) 

Standard at ResOx 5695-H

 

Application range 

Protection gases in general, 
precise measuring at low O

(e.g.  

<0,5 % Vol. O

2

) or  

concentrations  > 35 % Vol. O

2

   

Protection gases  with CO

2

 

concentrations and O

2

 

concentrations below < 35 % Vol. O

2

 

Permanent operation with high CO

conc. 

+++ 

Short time exposition to CO

2

  

+++ 

Use up to 100 % Vol O

2

 

+++ 

Use below 0.5% Vol O

2

 

+++ 

Speed /t

90

 

++ / <10 s 

++ / <10 s 

Life time 
/hours per % Vol O

2

, / at air

 

 /500.000 %h/ >2 years 

++ 

 /1.200.000 %h/ max. 6 years 

Measuring range 

  O

2

 partial pressure 

0 ... 1100 hPa  

0 ... 350 hPa 

  O

2

 concentration 

0.0 ... 100.0 % Vol. 

O

2

   

0.

0 … 35.0 % Vol. 

O

2

  

(reduced 

precision  above)

 

 

 

 

 

 

 

 

 

 