Operating Manual GMK 210
page 16 of 20
_____________________________________________________
_____________________________________________________________________________
11 Configuration of the device
Follow these instructions to configure the functions of the device:
Switch the device off.
Press
and keep it pressed, while turning the device on (press
shortly).
Release the sort-
button not before the first parameter “
P.OF
” is displayed.
Set parameter with up
or down
.
Jump to the next parameter by pressing
.
Parameter Value
Information
button
buttons
.
Detector Mode
factory setting: oF
OF
Device measures absolute humidity parameters (%)
– material curves are
available
ON
Device operates as a detector (relative measurement only) with adjustable
sensitivity
.
/
.
Detector Mode:
Sensitivity
factory setting: 10
5
…
100
Sensitivity
of the bar graph and acoustic signal
Set value corresponds to maximum rating via bar / acoustic signal
.
Auto Power-Off
(turn-off delay)
factory setting: 20 min
1
…
120
Auto Power-Off
(turn-off delay) in minutes. If no key is pressed for the time
adjusted here, the device is automatically switched off (adjustable 1 … 120 min)
OF
Auto power-off is deactivated (continuous operation)
.
Display unit
factory setting: %u
%u
Measuring value displayed in moisture %u
%w
Measuring value displayed in water content %w
.
Backlight
factory setting: 5
OF
No backlight
5
…
120
Backlight turns automatically off after 5 …120s (battery saving)
ON
Backlight always on as long as device is switched on
.
Acoustic signal
factory setting: on
OF
No acoustic signal
ON
Acoustic signal for moisture rating
.
Auto Hold
factory setting: oF
OF
Auto Hold deactivated: pressing hold-key freezes measuring and releases it
again
ON
Auto Hold activated:
hold-
key starts new measuring, the display will be „frozen‟
as soon as a stable measuring value was detected.
.
Restore factory settings
NO
Parameters are not changed to factory settings.
G
o
ATTENTION:
All parameters are changed to factory settings.
Press again to store changed settings, the device restarts (segment test).
NOTE:
If there is no key pressed within the menu mode within 2 minutes, the configuration will be
cancelled, the entered settings are lost!