background image

H68.0.02.6C-04 

 Operating Manual GMK 210 

page 4 of 20 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

2.2  Safety guidelines 

This device has been designed and tested in accordance with the safety regulations for electronic 
devices. However, its trouble-free operation and reliability cannot be guaranteed unless the 
standard safety measures and special safety advises given in this manual will be adhered to when 
using the device. 

1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not 

subjected to any other climatic conditions than those stated under "Specification". 
If the device is transported from a cold to a warm environment condensation may cause in a 
failure of the function. In such a case make sure the device temperature has adjusted to the 
ambient temperature before trying a new start-up. 

2.   

  If there is a risk whatsoever involved in running it, the device has to be switched off 

immediately and to be marked accordingly to avoid re-starting. 
Operator safety may be a risk if: 
- there is visible damage to the device 
- the device is not working as specified 
- the device has been stored under unsuitable conditions for a longer time. 
In case of doubt, please return device to manufacturer for repair or maintenance. 
 

3.  When connecting the device to other devices the connection has to be designed most 

thoroughly as internal connections in third-party devices (e.g. connection GND with protective 
earth) may lead to undesired voltage potentials that can lead to malfunctions or destroying of the 
device and the connected devices. 
 

This device must not be run with a defective or damages power supply unit.  
Danger to life due to electrical shock!  

 

4. 

 

 

Do not use these products as safety or emergency stop devices or in any other 
application where failure of the product could result in personal injury or material 
damage. 
Failure to comply with these instructions could result in death or serious injury and 
material damage. 

 

5. 

 

 

This device must not be used at potentially explosive areas! The usage of this 
device at potentially explosive areas increases danger of deflagration, explosion or 
fire due to sparking. 

 

Product Description 

3.1  Scope of supply 

The scope of supply includes:  

  GMK 210 
  9V - battery  
  Operating manual 
  Short manual 

3.2  Operation and maintenance advice 

If „bAt‟ is shown in the lower display the battery has been used up and needs to be replaced. 
However, the device will operate correctly for a certain time. 

The battery has to be taken out, when storing device above 50°C.  
We recommend taking out battery if device is not used for a longer period of time. 

 

 

 

 

 

 

 

 

Содержание GMK 210

Страница 1: ...info greisinger de as of version 1 0 GMK 210 Operation Manual Material moisture meter for non destructive measurements Application camper caravan and boat WEEE Reg Nr DE 93889386 JSP s r o Raisova 547...

Страница 2: ...MEDIUM DRY 6 5 3 HOLDING THE DEVICE 7 5 4 MOISTURE U AND WATER CONTENT W 7 5 5 CHARACTERISTICS AND MATERIALS 8 5 6 ZEROING FUNCTION 8 6 OPERATING MODES MEASURING MODE AND DETECTOR MODE 9 6 1 MEASURIN...

Страница 3: ...mmend the GMK 100 wood floor screed and plaster characteristics CM humidity The safety requirements see below have to be observed The device must be used only according to its intended purpose and und...

Страница 4: ...ns for a longer time In case of doubt please return device to manufacturer for repair or maintenance 3 When connecting the device to other devices the connection has to be designed most thoroughly as...

Страница 5: ...c Zeroing function see chapter 5 6 Key 2 and 3 together sort down Material selection downwards see chapter 5 5 Depth switch on the right side Sliding switch up ca 10 mm measuring depth d 10 Sliding sw...

Страница 6: ...rear side the area is marked under that it is predominantly measured For precise measurements the device has to lie on the measured material not only at the marked measuring spot but at its whole rear...

Страница 7: ...e u relating to oven dry mass Water content w is normally used for combustibles i e wood briquette You can choose whether the device should use moisture or water content see chapter 11 Moisture u rela...

Страница 8: ...ferent types of wood and their corresponding characteristics d xx in the addendum The materials are selected with these keys material selection upwards press Material selection downwards press at the...

Страница 9: ...reshold dEt SCL metal structures beams in plasterboard etc The measurement is relative only rEF curve Display in digit no unit After the detector mode is enabled Configuration dEt on a reasonable alar...

Страница 10: ...ensity dry for example a wood of density 0 68 kg dm3 is measured with material d 70 approximate actual density to the next material 0 52 d 50 0 53 d 55 8 Measurement of Other Materials The device has...

Страница 11: ...ea the device displays too high values in most cases Therefore metallic rods reinforcement aluminum insulation blankets metal wires etc can cause problems 9 1 Materials and design It is important to k...

Страница 12: ...his measurements cannot be performed However a relative measurement may be still possible therefore the displayed measuring value is compared to an value measured at a definitely dry spot of the same...

Страница 13: ...to preserve leak tightness warranty This is reasonable but with costs Defects can lead to massive humidity damage within only little time For example mildew can form within few month Therefore respon...

Страница 14: ...values Vacuum air 1 Polyester and Epoxy resin 3 4 Glass fiber 6 7 Water 80 ATTENTION Structures containing carbon fibers or structures with metallic graphitic coatings Metallic elements will also cau...

Страница 15: ...a long time This must not mandatory be problematic However if the measuring value exceeds 3 there should be further investigations Areas with pimples or water accumulations not seen from the outside c...

Страница 16: ...120 Auto Power Off turn off delay in minutes If no key is pressed for the time adjusted here the device is automatically switched off adjustable 1 120 min OF Auto power off is deactivated continuous o...

Страница 17: ...slope correction in Slope correction for measuring depth 25 mm factory setting of 0 OF 19 19 Value of slope correction in Press again to store changed settings the device restarts segment test NOTE I...

Страница 18: ...RH non condensing Storage temperature 25 to 70 C Power supply 9 V battery type IEC 6F22 included Power consumption measurement Approx 0 12 mA battery life time more than 2500 hours for alkaline batte...

Страница 19: ...flora d 50 Mahogany Swietenia spp d 50 Mahogany African Khaya spp d 50 Mahogany Philippine Parashorea plicata Shorea almon d 50 Maple New Guinea Flindersia pimentelianan d 55 Maple Sugar Acer saccharu...

Страница 20: ...H68 0 02 6C 04 GREISINGER electronic GmbH D 93128 Regenstauf Hans Sachs Stra e 26 Tel 09402 9383 0 Fax 09402 9383 33 eMail info greisinger de...

Отзывы: