5 Start Operation
Connect sensors, turn device on via
key.
After segment test
the device displays some configuration:
if standard measurement is selected
if fast measurement is selected
if peak value detection is selected
After that the device is ready for measuring.
6 Configuration
Some menu points depend on current device settings.
To change device settings, press
“menu”
for 2 seconds. This will activate the configuration menu (main
display: “SEt”). Pressing
“menu”
changes between the menus points, pressing
jumps to the
referring parameters, which can be selected with key
.
The parameters can be changed with
or
.
Pressing
“menu”
again jumps back to the main configuration menu and saves the settings.
“enter”
finishes the configuration and returns to standard measuring operation.
Pressing “menu” and “store” at the same time for more than 2 seconds will reset the device
to factory defaults.
If no key is pressed for more than 2 minutes the configuration will be aborted. All changes
will not be saved!
Menu
Parameter
Value
Description
or
Set Configuration: General configurations
Unit: Select measuring unit
**
Arrow to
bar,
mbar …
Measuring value is displayed in corresponding unit, the selectable units
depend on connected sensor
Arrow to
“user”
Measuring value is displayed in freely adjustable user-unit
bar, mbar
Unit “user”: base unit
0000, 000.0,
… .0000“
Unit “user”: decimal point setting (only if user-unit selected)
-19999...19999
Unit “user”: multiplication factor
Sea Level: Sea-level correction
**
oFF
Sea-level correction off
on
Sea-level correction on
(p.r.t. 7.4)
-2000 ... 9999
Sea-level in [m]
**
Rate: Measuring rate
Slo
Slow
measuring rate (4 Hz filtered, low power consumption)
FASt
Fast
measuring rate, filtered (1000 Hz)
P.dEt
Peak detection
: fast measuring rate, unfiltered (1000 Hz)