 Operating Manual GMH 5130 

page 10 of 14 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

8  Output 

The output can be used as serial interface (for USB 5100 interface converter).  
If the output is not needed, it is strongly recommended to deactivate it (Out oFF) to lower power 
consumption. This increases battery life time.  
If the device is used together with interface adapter USB 5100 the device is supplied from the interface. 

Pin assignment: 

 

4: external 5V, 50 mA 

3: GND 

2: TxD/RxD (3.3V Logic) 

1: not used  

 

 

 

Only suitable adaptor cables are 
permitted (accessories)!

 

 

 
 

8.1  Interface 

The device can be connected to a USB interface of a PC by the electrically isolated interface converter 
USB 5100 (accessory). The data is transmitted binary-coded and protected against transmission errors by 
complex safety mechanism (CRC).  

The following standard software packages are available: 

 

GSOFT3050

:  

Operating and evaluation software for the integrated logger function 

 

EBS20M / -60M

20-/60-channel software for measuring value display 

 

EASYControl net

: Universal multi-channel software for real-time recording and displaying of measuring 

 

data; with real database 

In case you want to develop your own software we offer a 

GMH3000-development package 

including: 

  a universally applicable Windows functions library ('GMH3x32e.DLL') with documentation, can be used 

by all ‘established’ programming languages, suitable for:

  

Windows 2000™, Windows XP™, Windows Vista™, Windows 7™ 

  Programming examples

 

Visual Basic 4.0™, Delphi 1.0™, Testpoint™ etc. 

The device has 3 channels: 

-

 

Channel 1: current measuring value (sensor 1) and base address 

-

 

Channel 2: min peak   

 

-

 

Channel 3: max peak  

 

 

 

The measuring-/ alarm- and display range values read back from the interface are always in 
the selected measurement unit (mbar, bar...)! 

 

9  Input Adjustment 

The zero point and slope of each measuring inputs can be adjusted with the parameters offset (“OFFS”) and 
scale (“SCAL”).   
A reasonable adjustment presumes reliable references (e.g. ice water, controlled precision water bath, etc.). 

If the inputs are adjusted (i.e. offset and scale are different from default settings) the device will shortly 
display “Corr” after turned on.  

Default setting for offse

t and scale are ‘off’ = 0.0, i.e. inputs are not changed.  

Zero point correction: 

Displayed value = measured value 

– OFFS  

Zero point and slope correction: 

Displayed value = (measured value 

– OFFS) * (1 + SCAL / 100 )    

 

 

Содержание GMH 5130

Страница 1: ...49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de WEEE Reg Nr DE 93889386 as of version 1 0 GMH 5130 Operating Manual Handheld Pressure Meter water proof Please carefully read these instructions b...

Страница 2: ...LEMENTS 5 4 2 PUSHBUTTONS 5 4 3 CONNECTIONS 5 4 4 POP UP CLIP 6 5 START OPERATION 7 6 CONFIGURATION 7 7 REMARKS TO SPECIAL FEATURES 8 7 1 DIFFERENT PRESSURE UNITS 8 7 2 DIFFERENT KINDS OF MEASURING 8...

Страница 3: ...to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Prot...

Страница 4: ...e usage of this device at potentially explosive areas increases danger of deflagration explosion or fire due to sparking 3 Product Specification 3 1 Scope of supply The scope of supply includes GMH 51...

Страница 5: ...ase refer to chapter 7 1 7 corr arrow zero point or slope correction is active 4 2 Pushbuttons On off key backlight press shortly activate backlight or switch on instrument press longer switch off ins...

Страница 6: ...to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attac...

Страница 7: ...es the settings enter finishes the configuration and returns to standard measuring operation Pressing menu and store at the same time for more than 2 seconds will reset the device to factory defaults...

Страница 8: ...sor dep e g 5 00 5 00 mbar The offset of sensor will be displaced by this value to compensate for deviations of the probe or of the measuring device Slope adjustment of sensors oFF No slope adjustment...

Страница 9: ...ing tube connection for lower press in contact to ambient air by means of a tube no water contact and bring the sensor with its open pressure connection for higher pressure to water depth to be measur...

Страница 10: ...l net Universal multi channel software for real time recording and displaying of measuring data with real database In case you want to develop your own software we offer a GMH3000 development package...

Страница 11: ...o G1 2 pressure terminal or plug plastic tube to a suitable adapter Connection to instrument Use cable MSD K51 11 Accuracy Check Adjustment Service You can send the device to the manufacturer for adju...

Страница 12: ...h to be displayed Err 4 Display range underflow Check value below 19999 Tara to low Er 11 Value could not be calculated Choose different unit Calculation overflow happened Choose different unit Err 7...

Страница 13: ...ment display illuminated white Correction function offset slope via menu if offset slope value not equal zero corr is displayed Additional functions min max hold function Averaging adjustable 1 120 se...

Страница 14: ...H82 0 01 6C 02 Operating Manual GMH 5130 page 14 of 14 _____________________________________________________ _____________________________________________________________________________...

Отзывы: