GREISINGER GMH 3830 Скачать руководство пользователя страница 6

H60.0.11.6C-16 

 Operating Manual GMH 3830 

Page 6 of 18 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

4  Device Configuration 

 

For configuration of the device press "

Menu

"-key (key 4) for 2 seconds, the first menu will be shown. 

Choose between the individual values that can be set by pressing the "

Menu

"-key (key 4) again. 

The individual values are changed by pressing the keys  "

" (key 2) or "

" (key 5). 

Use  key "

Store/

" (key 6) to leave configuration and to store settings. 

 

Parameter 

Values 

Meaning

 

 

key 

Menu 

key

 

or

 

 

 

p.r.t. 

Sort: limitation of the material selection

 

 

 

off

Unrestricted material selection via key 2 and 5 

 

1...8

:

 

Material selection in-between 1 up to 8 selectable materials 

....

 

selectable materials (not available if Sort = off) 
Select the desired material that should be available during 
the measure via key 2 and 5. 

 

Generic Settings

 

 

 

Arrow bottom left 
points to

 

“%u”

 

Moisture display = moisture content [%u] 

 

Arrow bottom left 
points to

 

“%w” 

Moisture display = wet-basis moisture content [%u] 

 

°C 

All temperature values are in degrees Celsius 

 

°F 

All temperature values are in degrees Fahrenheit 

 

oFF

 

Atc off: temperature input for compensation via keys 

5.4 

on 

Atc on: temperature compensation via internally measured temperature 
or external probe 

 

oFF

 

Auto HLD off: continuous measuring. 

5.3 

on 

Auto-HLD on: when reaching a stable measuring result, this will be 
frozen with-HLD. When pressing the store-key a new measure will be 
initiated. If logger is switched on (‚Func CYCL‘, ‚Func Stor‘): device 
works like setting would be “auto-HLD off”

 

 

1...120 

Power-off delay in minutes.  
Device will be automatically switched off as soon as this time has 
elapsed if no key is pressed/no interface communication takes place 

 

oFF 

Power-off function inactive (continuous operation, e.g. mains operation) 

 

oFF 

Function of the output: No output function, lowest power consumption 

 

SEr 

Output is serial interface 

dAC 

Output is analogue output 0...1V 

 

01, 11 ... 91 

Base Address when Output = Serial Interface : 
 Base address of device for interface communication. 

7.1 

 

0.0 ... 100.0% 

Enter desired moisture value at which the analogue output potential 
should be 0V 

7.2 

 

0.0 ... 100.0% 

Enter desired moisture value at which the analogue output potential 
should be 1V 

7.2 

 

 

The settings will be set to the settings ex works, if keys ‘Set‘ and ‘Store‘ are pressed  

 simultaneously for more than 2 seconds. 

 

 

Содержание GMH 3830

Страница 1: ...3128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de as of version 2 1 Operating Manual GMH 3830 Resistive material moisture and temperature measuring Instrument keep for future refer...

Страница 2: ...G OTHER MATERIALS 8 5 6 1 Hard Materials concrete or similar Measuring with brush type probes GBSL91 or GBSK91 8 5 6 2 Soft Materials polystyrene or similar Measuring with Measuring rods or pins GMS 3...

Страница 3: ...n indirect impact on the operation or can trigger an unforeseen reaction if ignored table 1 2 2 Safety Instructions This device has been designed and tested in accordance to the safety regulations for...

Страница 4: ...ove specification do not throw hit against etc Protect plugs and sockets from soiling 4 Mains Operation When using a power supply unit please note that operating voltage has to be 10 5 to 12 V DC Do n...

Страница 5: ...splay elements 3 Moisture evaluation Evaluation of the material condition via top arrows DRY MEDIUM WET 4 Warning triangle Indicates low battery 5 u or w Displays unit moisture content u or wet basis...

Страница 6: ...F All temperature values are in degrees Fahrenheit oFF Atc off temperature input for compensation via keys 5 4 on Atc on temperature compensation via internally measured temperature or external probe...

Страница 7: ...re measuring devices use one single group for spruce and oak in reality the deviation of these characteristic curves is more than 3 Base for this statement are complex statistical surveys considered m...

Страница 8: ...ure dispersion may be irregular often in the core is more moisture than on the edge Surface moisture The wood edge could be more humid than the core if the wood had been stored outside and e g was in...

Страница 9: ...measuring value an individual moisture estimation will be displayed simultaneously This moisture evaluation is only a guidance value the final evaluation is depending on the application of the materi...

Страница 10: ...tion x X 0 read nominal value x x 202 read unit of display x X 3 read system status x x 204 read decimal point of display x 12 read ID no x 205 read extended measuring type in display x X 176 read min...

Страница 11: ...o valid signal Connect meas material charge at the probe device will discharge esp at dry wood Wait until probe has discharged Sensor broken or device defective return to manufacturer for repair Err 1...

Страница 12: ...g Output 3 5mm audio plug stereo Selectable as serial interface via optically isolated interface adapter GRS3100 GRS3105 or USB3100 p r t accessories directly connectable to RS232 or USB interfaces or...

Страница 13: ...Japanese Fraxinus mandshurica h 134 4 100 Ash Red Flindersia excelsa h 21 5 86 Ash Scaly Ganophyllum falcatum h 22 5 100 Ash Silver Northern Flindersia schottina h 23 7 89 Ash Silver Queensland Flinde...

Страница 14: ...ustralasicum h 113 7 85 Cupiuba Goupia glabra h 147 6 69 Curupix Micropholis h 114 6 63 Cypress Cupressus spp h 456 5 100 Cypress Northern Callitris intratropica h 115 6 100 Cypress Rottnest Island Ca...

Страница 15: ...Mahogany White Eucalyptus acmenoides h 247 6 100 Mahogony Khaya Khaya spp h 235 7 100 Mahogony American Swietenia spp h 234 6 100 Mahogony Phillipines Parashorea plicata h 236 5 100 Mahogony Phillipi...