12 Specification
Measuring
Channel1
Channel2
-
Principle
Resistive material-moisture-measuring
Temperature-measuring thermocouple type K
matching DIN EN 13183-2: 2002
or internal temperature-measuring
-
Char. curves
466 different kinds of wood
matching DIN EN 60584-1: 1996, ITS90
28 different building materials
-
Probe connection
BNC Plug
floating connector for mini-blade-terminal
-
Display range
0.0...100.0 % moisture content
thermocouple: -40.0... +200.0°C / -40.0... + 392.0°F
(depending on characteristic curve)
int. temp.-Meas.: -30.0...75.0°C / -22.0...167.0°F
equal to ca. 3kOhm ... 2TerraOhm
-
Specified measuring range: Wood
8 … 40% u, Wood temperature 0 … 40°C (no frozen wood!)
-
Resolution
0,1% moisture content
0.1°C / 0,1°F
-
Evaluation
Evaluation of the material condition in 9 steps from DRY to WET
-
Accuracy Device without probe
±1Digit (at nominal-temperature)
Wood: ±0.2% moisture content (deviation from
Type K: ± 0.5% m.v. ± 0.3°C
characteristic curve, range 6..40%)
int. t.-measuring: ± 0.3°C (is type K reference junction)
building mat.: : ±0.2% moisture content (dev. from
char. curve, range depending on char. curve)
-
Total accuracy
refer to chapter
„11 Measuring precision“
-
Temperature drift
< 0.005 % moisture content per 1K
0.01% per 1K
Nominal temperature
25°C
Ambient
Temperature
-
25 ... +50°C (-13 .. 122°F)
Relative humidity 0 ... 95 %RH (non condensing)
Storage temperature -
25 ... +70°C (-13 ... 158°F)
Housing
Dimension: 142 x 71 x 26 mm (L x B x D)
impact resistant ABS, membrane keyboard, transparent panel.
Front side IP65, integrated pop-up-clip for table top or suspended use
Weight
approx. 155 g
Output:
3.5mm audio plug, stereo
Selectable as serial interface:
via optically isolated interface adapter GRS3100, GRS3105 or USB3100 (p.r.t. accessories)
directly connectable to RS232- or USB-interfaces.
or analogue output:
0..1V, freely scalable (resolution 13bit, accuracy 0.05% at nominal temperature, cap. load <1nF)
Power Supply
9V-Battery, type IEC 6F22 (included) as well as additional d.c. connector (diameter of internal pin 1.9 mm) for
external 10.5-12V direct voltage supply. (suitable power supply: GNG10/3000)
Power Consumption
output off
approx. 2.5mA
output serial interface:
approx. 2.7mA
analogue output:
approx. 3.0mA
cyclic logger sleeping state with output deactivated:
< 0.1mA
cyclic logger sleeping state with activated serial interface: < 0.3mA
Display
Two 4 digits LCD’s (12.4mm high and 7 mm high) for material moisture temperature or characteristic curve,
hold function, etc. as well as additional pointing arrows.
Pushbuttons
6 membrane keys for on/off switch, menu operation, characteristic curve, hold-function etc.
Hold Function
Press button to store current value.
Automatic-Off-Function
Device will be automatically switched off if no key is pressed/no interface communication takes place for
the time of the power-off delay. The power-off delay can be set to values between 1 and 120 min.; it can be
completely deactivated.
EMC:
The device corresponds to the essential protection ratings established in the Directives of the European
Parliament and of the council on the approximation of the laws of the member states relating to the
electromagnetic compatibility (2004/108/EC).
EA1 +A2 (Appendix B, class B), additional error: < 1% FS
13 Disposal Notes
Dispense exhausted batteries at destined gathering places.
This device must not be disposed as “residual waste”. To dispose this device, please send it
directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.