H49.0.01.6C-03
Operating Manual GMH 3710
Page 6
4.4.2 Analogue Output – Scaling with DAC.0 and DAC.1
With the DAC.0 and DAC.1 values the output can be rapidly scaled to Your efforts.
Keep in mind not to connect low-resistive loads to the output, otherwise the output value will be wrong and battery life is
decreased. Loads above ca 10kOhm are uncritical.
If the display exceeds the value set by DAC.1, then the device will apply 1V to the output
If the display falls below the value set by DAC.0, then the device will apply 0V to the output
In case of an error (Err.1, Err.2, no sensor, etc.) the device will apply slightly above 1V to the output.
GND
plug wiring::
+Uout
Attention!
The 3
rd
contact has to be left floating!
Only stereo plugs are allowed!
5 Fault and System Messages
Display
Meaning
Remedy
low battery voltage, device will continue to work for a short
time
replace battery
If mains operation: wrong voltage
replace power supply, if fault continues to
exist: device damaged
low battery voltage
replace battery
If mains operation: wrong voltage
Check/replace power supply, if fault continues
to exist: device damaged
low battery voltage
replace battery
If mains operation: wrong voltage
Check/replace power supply, if fault continues
to exist: device damaged
system error
Disconnect battery or power supply, wait
some time, re-connect
No display
or
weird display
Device does not
react on keypress
device defective
return to manufacturer for repair
Sensor error: no sensor connected
Connect sensor to socket
----
sensor/cable or device defective
return to manufacturer for repair
Value exceeding measuring range
Check: Is the value exceeding the measuring
range specified? ->temperature too high!
Wrong probe connected
Check probe
Err.1
sensor/cable defective
-> replace
Value below display range
Check: Is the
value below the measuring
range specified? -> temperature too low
!
Wrong probe connected
Check probe
Err.2
sensor/cable defective
-> replace
Err.3
Value exceeding display range
-> set resolution to 0.1° or Auto
Err.4
Value below display range
-> set resolution to 0.1° or Auto
Err.7
system error
return to manufacturer for repair
6 Calibration Services
Calibration certificates can be issued by the factory. For this the device has to be sent to the manufacturer. Just the
manufacturer can check the factory settings and correct them if necessary.