Operating Manual GMH 3710

Page 6

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

4.4.2  Analogue Output – Scaling with DAC.0 and DAC.1

With the DAC.0 and DAC.1 values the output can be rapidly scaled to Your efforts.
Keep in mind not to connect low-resistive loads to the output, otherwise the output value will be wrong and battery life is
decreased. Loads above ca 10kOhm are uncritical.
If the display exceeds the value set by DAC.1, then the device will apply 1V to the output
If the display falls below the value set by DAC.0, then the device will apply 0V to the output
In case of an error (Err.1, Err.2, no sensor, etc.) the device will apply slightly above 1V to the output.

GND

plug wiring::

+Uout

Attention!

The 3

rd

 contact  has to be left floating!

Only stereo plugs are allowed!

5  Fault and System Messages

Display

Meaning

Remedy

low battery voltage, device will continue to work for a short
time

replace battery

If mains operation: wrong voltage

replace power supply, if fault continues to
exist: device damaged

low battery voltage

replace battery

If mains operation: wrong voltage

Check/replace power supply, if fault continues
to exist: device damaged

low battery voltage

replace battery

If mains operation: wrong voltage

Check/replace power supply, if fault continues
to exist: device damaged

system error

Disconnect battery or power supply, wait
some time, re-connect

No display

or

weird display

Device does not

react on keypress

device defective

return to manufacturer for repair

Sensor error: no sensor connected

Connect sensor to socket

----

sensor/cable or device defective

return to manufacturer for repair

Value exceeding measuring range

Check: Is the value exceeding the measuring
range specified? ->temperature too high!

Wrong probe connected

Check probe

Err.1

sensor/cable defective

-> replace

Value below display range

Check: Is the 

value below the measuring

range specified? -> temperature too low

!

Wrong probe connected

Check probe

Err.2

sensor/cable defective

-> replace

Err.3

Value exceeding display range

-> set resolution to 0.1° or Auto

Err.4

Value below display range

-> set resolution to 0.1° or Auto

Err.7

system error

return to manufacturer for repair

6 Calibration Services

Calibration certificates can be issued by the factory. For this the device has to be sent to the manufacturer. Just the
manufacturer can check the factory settings and correct them if necessary.

Содержание GMH 3710

Страница 1: ...free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under Specification 2 If the device is transported from a...

Страница 2: ...e probe the plug will slide in smoothly if plug is entered correctly Selection of Output Mode The output can be used as serial interface or as analogue output This choice has to be done in the configu...

Страница 3: ...measuring will be changed by this factor to compensate for deviations in the temperature probe or in the measuring device factor is in oFF Scale correction factor inactive 0 000 2 5 Power off Selecti...

Страница 4: ...ate very much from the ambient temperature measuring errors often occur if the heat loss caused by the probe is not considered When measuring fluids therefore the probe should be emerged sufficiently...

Страница 5: ...e device can be connected to a PC With the GRS3105 it is possible to connect up to 5 instruments to a single interface please also refer to GRS3105 manual As a precondition the base addresses of all d...

Страница 6: ...r supply if fault continues to exist device damaged low battery voltage replace battery If mains operation wrong voltage Check replace power supply if fault continues to exist device damaged system er...

Страница 7: ...S232 interfaces or analogue output 0 1V freely scaleable resolution 13bit accuracy 0 05 at nominal temperature cap load 1nF Power Supply 9V Battery type IEC 6F22 included as well as additional d c con...

Страница 8: ...H49 0 01 6C 03 Operating Manual GMH 3710 Page 8...

Отзывы: