2 Device Configuration

For configuration of the device press "

Menu

"-key (key 4) for 2 seconds, the first menu will be shown.

Choose between the individual values that can be set by pressing the "

Menu

"-key (key 4) again.

The individual values are changed by pressing the keys  "

5

" (key 2) or "

6

" (key 5).

Use  key "

Quit

" (key 6) to leave configuration and to store settings.

2.1  'Unit': Selection of Temperature Unit °C /°F

°C

:

All temperature values are in degrees Celsius

°F

:

All temperature values are in degrees Fahrenheit

2.2  'Resolution': The Display Resolution

0.1°

:

Resolution 0.1°C

0.01°

Resolution 0.01°C

Auto

:

Resolution is selected automatically

2.3  'Offset': Zero Displacement

-2.50°C...2.50°C
   respectively
-4.50°F...4.50°F

The zero point of the measurement will be displaced by this value to
compensate for deviations in the temperature probe or in the
measuring device

oFF

:

Zero displacement inactive (=0.0°)

2.4  'Scal': Scale Correction

-2.000...2.000:

The scale of the measuring will be changed by this factor to
compensate for deviations in the temperature probe or in the
measuring device (factor is in %)

oFF

:

Scale correction factor inactive (=0.000)

2.5  'Power.off': Selection of Power-Off Delay

1...120

:

Power-off delay in minutes.
Device will be automatically switched off as soon as this time has elapsed if
no key is pressed/no interface communication takes place

oFF

:

Power-off function inactive (continuous operation, e.g. mains operation)

2.6  'Out': Function of the Output

OFF:

No output function, lowest power consumption

SEr:

 

Output is serial interface

dAC:

Output is analogue output 0...1V

2.7  'Address': Selection of Base Address when Output = Serial Interface

01, 11, 21, ..., 91

:

Base address of device for interface communication.

2.8  'dAC.0Volt': Output Offset When Output = Analogue Output

-200.0...850.0°C
    respectively
-328.0...1562.0°F

Input of the temperature at which 0V should be output

2.9  'dAC.0Volt': Output Scale When Output = Analogue Output

-200.0...850.0°C
    respectively
-328.0...1562.0°F

Input of the temperature at which 1V should be output

Hint:  If keys ‘Set‘ and ‘Store‘ are pressed simultaneously for more than 2 seconds, the settings are reset to

the settings ex works

Содержание GMH 3710

Страница 1: ...free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under Specification 2 If the device is transported from a...

Страница 2: ...e probe the plug will slide in smoothly if plug is entered correctly Selection of Output Mode The output can be used as serial interface or as analogue output This choice has to be done in the configu...

Страница 3: ...measuring will be changed by this factor to compensate for deviations in the temperature probe or in the measuring device factor is in oFF Scale correction factor inactive 0 000 2 5 Power off Selecti...

Страница 4: ...ate very much from the ambient temperature measuring errors often occur if the heat loss caused by the probe is not considered When measuring fluids therefore the probe should be emerged sufficiently...

Страница 5: ...e device can be connected to a PC With the GRS3105 it is possible to connect up to 5 instruments to a single interface please also refer to GRS3105 manual As a precondition the base addresses of all d...

Страница 6: ...r supply if fault continues to exist device damaged low battery voltage replace battery If mains operation wrong voltage Check replace power supply if fault continues to exist device damaged system er...

Страница 7: ...S232 interfaces or analogue output 0 1V freely scaleable resolution 13bit accuracy 0 05 at nominal temperature cap load 1nF Power Supply 9V Battery type IEC 6F22 included as well as additional d c con...

Страница 8: ...H49 0 01 6C 03 Operating Manual GMH 3710 Page 8...

Отзывы: