GREISINGER GMH 3151 Скачать руководство пользователя страница 6

H61.0.02.6C-04

User’s Manual GMH3151

page 6 of 12

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         

_______________________________________________________________________________________________________________________

GREISINGER electronic GmbH

2.5  Zero Displacement Sensor ('OFFS')

A zero displacement can be carried out for the measured value:

value displayed = value measured  -  offset

Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried out. Together with the scale correction
(see below) this factor is mainly used to compensate for sensor deviations. Input is in the display unit.

2.6  Scale Correction Sensor ('SCAL')

The scale of the measuring can be influenced by this setting (factor is in %):

displayed value = measured value * (1+Scal/100)

Standard setting: 'off' =0.000, i.e. value is not corrected. Together with the zero displacement (see above)
this factor is mainly used to compensate for sensor deviations.

2.7  Power Off Time

If there won’t be pressed any key and no interface communication takes place for the time of the power off
time setting (P.Off), the device will be switched off automatically to save battery power.
If P.oFF = oFF then the automatic switch off is deactivated.

2.8 Output

The output can be used as serial interface (for GRS3100 or GRS3105 interface adapters) or as analogue
output (0-1V).

2.8.1  Interface - Base Address ('Adr.')

Up to 10 devices of the GMH3xxx- handheld-family can be connected to a serial interface at once
(depending on interface converter, e.g. GRS3105: 5 devices). To get access to each device the base
addresses of the devices have to be different. For example choose 01 for the first, 11 for the second device
and so on. See also chapter 4.

2.8.2  Analogue Output – Scaling with DAC.0 and DAC.1

Note: Analogue output can not be used during logger recordings

With the DAC.0 and DAC.1 values the output can be rapidly scaled to Your efforts.
Keep in mind not to connect low-resistive loads to the output, otherwise the output value will be wrong and
battery life is decreased. Loads above ca 10kOhm are uncritical.
If the display exceeds the value set by DAC.1, then the device will apply 1V to the output
If the display falls below the value set by DAC.0, then the device will apply 0V to the output

In case of an error (Err.1, Err.2, no sensor, etc.) the device will apply slightly above 1V to the output.

GND

plug wiring::

+Uout

Attention!

The 3

rd

 contact  has to be left floating!

Only stereo plugs are allowed!

2.9 Alarm

3 possible settings: Alarm off (AL.oFF), on with horn sound (AL.on), on without horn (AL.no.So).
Following conditions will display an alarm, when the function is activated (on or no.So):

-

  Value is below lower (AL. Lo) or above upper alarm rail (AL.Hi).

-

  Sensor error (Sens Erro)

-

  Low battery (bAt)

-

  Fe 7: System error (always with sound)

In case of an alarm,  and when polling the interface the prio-flag is set in the returned interface message.

2.10 Real Time Clock

The real time clock is used for the logger function: Recorded values are also containing the point of time,
when they were measured. Please check the settings when necessary.
If the battery was replaced the referring menu ‚CLOC‘ will automatically be started.

Содержание GMH 3151

Страница 1: ...ng 5 2 2 MEASURING OF WATER LEVEL DISPLAY UNIT M 5 2 3 SEA LEVEL CORRECTION FOR ABSOLUTE PRESSURE SENSORS 5 2 4 AVERAGING FUNCTION 5 2 5 ZERO DISPLACEMENT SENSOR OFFS 6 2 6 SCALE CORRECTION SENSOR SCA...

Страница 2: ...ions for a longer period of time In case of doubt please return device to manufacturer for repair or maintenance 6 Warning Do not use these product as safety or emergency stop device or in any other a...

Страница 3: ...3 Set Menu Calling of configuration Tare Function By pressing Tara key 3 the displays will be set to 0 All measurings from then on will be displayed relatively to the set tare values When the tare fun...

Страница 4: ...Base address of interface if Out SEr dAC 0 eg 5 00 5 00 mbar Enter desired value at which the analogue output potential should be 0V if Out dAC SEt ConF dAC 1 eg 5 00 5 00 mbar Enter desired value at...

Страница 5: ...press in contact to ambient air by means of a tube no water contact and bring the sensor with its open press connection for higher pressure to water depth to be measured display and is compensated fo...

Страница 6: ...ces To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 4 2 8 2 Analogue Output...

Страница 7: ...key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is displayed see b...

Страница 8: ...the logger operation will be called The display will show By pressing Store again the recording will be initiated After that the display shows St XXXX for a short time whenever a measuring is recorded...

Страница 9: ...he interface activate the alarm function of the GMH and the relays output is operating The device has 3 channels Channel 1 current measuring value base address Channel 2 min peak p r t chapter 3 Chann...

Страница 10: ...E For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 6 Error And System Messages Display Meaning What to do Low batte...

Страница 11: ...he min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and year Logger 2 Functions individual value logger Func Stor and cyclic l...

Отзывы: