1.3 Connections

1. Output:

 Operation as interface: Connect to optically isolated

interface adapter (accessory: GRS 3100,  GRS3105 or USB3100)
Operation as analogue output: Connection via suitable cable.
Attention: The output mode has to be configured (p.r.t. 2.8) and
influences battery life!

2. Connection

 for 

pressure sensors

 of the GMSD and GMXD-family

3.

The 

mains socket

 is located at the left side of the instrument

1.4 Display

Units:

 an arrow points to the

chosen measuring unit

SL:

 appears if sea-level-

correction is activated

Tara:

 appears if tara-function

is activated

main display: 

shows measuring value

secondary display:

shows min-, max- or hold value

Logg:

appears, if a logger function is chosen

1.5 Basic Operation

When switching on 

the device and the logger function is not off the time of the integrated clock will shortly

be displayed. Furthermore the kind of measuring is displayed (Slo/Fast/P.det, p.r.t 2.1) and "Corr", if a
Sensor with activated offset or scale correction is connected.

After changing the battery

 the clock-setting menu is activated automatically (‚CLOC‘). Check the clock and

adjust, if necessary (p.r.t. chapter 2).

On-/Off-Switch
Tara:

Calling of tara function

min/max:

Showing the min- resp. max-memory

Store/Quit:

  Calling of hold function resp. calling of logger functions (p.r.t. chapter 3)

Set/Menu:

Calling of configuration

Tare Function:

By pressing ´Tara´ (key 3) the displays will be set to 0. All measurings from then on will be
displayed relatively to the set  tare values. When the tare function is activated, the arrow
"Tara" appears in the display. To deactivate tare function press ´Tara´  for >2 seconds.

Please Note: Activating/deactivating tara clears the max- & min-memories.

Max Memory:

Pressing ´max´ (key 2) shows the maximum of the measured value. Pressing it again
hides them. To clear the max. memory press key ´max´ for >2 seconds.

Min Memory:

Pressing ´min´ (key 5) shows the minimum of the measured value. Pressing it again hides
them. To clear the min. memory press key ´min´ for >2 seconds.

Hold Function:

By pressing ´Store/Quit´ (key 6) the last measuring value will be held in the secondary
display. Pressing it again hides it. (only when logger = ‚off‘).

Zero-Point Adjustment: 

If there is no pressure or zero-pressure (absolute) applied to the pressure ports the

device will display 0. If  there is a permanent deviation (and device is operated under
steady conditions), a permanent zero point adjustment can be carried out. To carry out the
adjustment press button 3 for approx. 5 seconds (Auto Null will be displayed shortly). The
adjustment is done via the OFFSET-value of the sensor (referring configuration menu).
Please note: A zero-point adjustment can only be carried out if  the difference between the
value on display is less than 500 digits!
To recall the manufacturer’s calibration press button 3 for approx. 15 seconds.

Note: If a zero-point adjustment was carried out, this will be signalled by the short
         displaying of „Corr“ when switching on the device.

1

2

3

4

5

6

ON

OFF

max

Tara

Store

min

Set

Menu

Quit

Содержание GMH 3151

Страница 1: ...ng 5 2 2 MEASURING OF WATER LEVEL DISPLAY UNIT M 5 2 3 SEA LEVEL CORRECTION FOR ABSOLUTE PRESSURE SENSORS 5 2 4 AVERAGING FUNCTION 5 2 5 ZERO DISPLACEMENT SENSOR OFFS 6 2 6 SCALE CORRECTION SENSOR SCA...

Страница 2: ...ions for a longer period of time In case of doubt please return device to manufacturer for repair or maintenance 6 Warning Do not use these product as safety or emergency stop device or in any other a...

Страница 3: ...3 Set Menu Calling of configuration Tare Function By pressing Tara key 3 the displays will be set to 0 All measurings from then on will be displayed relatively to the set tare values When the tare fun...

Страница 4: ...Base address of interface if Out SEr dAC 0 eg 5 00 5 00 mbar Enter desired value at which the analogue output potential should be 0V if Out dAC SEt ConF dAC 1 eg 5 00 5 00 mbar Enter desired value at...

Страница 5: ...press in contact to ambient air by means of a tube no water contact and bring the sensor with its open press connection for higher pressure to water depth to be measured display and is compensated fo...

Страница 6: ...ces To get access to each device the base addresses of the devices have to be different For example choose 01 for the first 11 for the second device and so on See also chapter 4 2 8 2 Analogue Output...

Страница 7: ...key 6 is pressed a measurement and its time stamp will be recorded The recorded data can be viewed either in the display when calling the configuration an additional menu REAd LoGG is displayed see b...

Страница 8: ...the logger operation will be called The display will show By pressing Store again the recording will be initiated After that the display shows St XXXX for a short time whenever a measuring is recorded...

Страница 9: ...he interface activate the alarm function of the GMH and the relays output is operating The device has 3 channels Channel 1 current measuring value base address Channel 2 min peak p r t chapter 3 Chann...

Страница 10: ...E For measurements of over under or absolute pressure screw sensor to G1 4 pressure terminal or plug plastic tube to a suitable adapter 6 Error And System Messages Display Meaning What to do Low batte...

Страница 11: ...he min and max rails set Alarming is done by integrated horn display and interface Real time clock Integrated clock with date and year Logger 2 Functions individual value logger Func Stor and cyclic l...

Отзывы: