GREISINGER GE 014 Скачать руководство пользователя страница 7

Z03.0.0X.6C-11 

 Operating manual GE xxx serie 

page 7 of 12 

 

 

 

 

 

The diaphragm, which establishes a connection between the electrolyte and the liquid to be 
measured, can be designed in different ways. Clogging or soiling of the diaphragm is a frequent 
cause of a malfunctioning or sluggish electrode. 

Always handle the glass membrane with extreme care. The so-called source layer forms there. This 
is crucial for the measurement and must always be kept moist. 

3.3  Further information 

A pH electrode is a wear part. If the signal is very slow or the required values are no longer 
observed after careful cleaning and possible regeneration, the electrode must be replaced. When 
using the electrodes, be aware that various substances in aqueous solutions can corrode glass and 
that chemicals can produce a chemical reaction with the KCl solution in the electrode, which can 
result in blockage of the diaphragm. 

Examples:  

•  In solutions that contain proteins, such as for measurements in medical and biological 

applications, KCl can cause denaturation of the protein. 

•  Coagulated paints 
•  Solutions that contain high concentrations of silver ions 

Substances that accumulate on the glass membrane or the diaphragm affect the measurement and 
must be removed regularly. This can be achieved for example with automatic cleaning systems. 

3.4  Service life 

 

The service life of electrodes is normally at least 8 to 10 months. When cared for 
properly, this can usually increase to more than 2 years. The actual life will vary 
depending on the particular application. 

Extreme pH values and high temperatures accelerate the aging process. 

 

Operating and Maintenance 

4.1  Measuring and storing 

The pH-electrodes have been tested and have been subordinated strict quality controls in all 
manufacturing-steps 
To keep the optimum efficiency and accuracy for a long time take care of the following points: 

  Remove the 

storing

 

protection-cap

 and rinse the shaft and pH-glass-diaphragm with distilled 

water. 

 

Important!

 The pH-glass-diaphragm has to be kept wet. When not in use the electrode must be 

stored in 3 mol/l KCl solution. (GE 103 in 1 mol/l KNO

3

). If the pH-glass-diaphragm dried out, 

the performance and the responsiveness are affected. To re-moisten the electrode store it in 3 
mol/l KCl solution for 24 hours (GE 103 1 mol/l KNO

3

). 

A longer storage of a combination-electrode or a reference-electrode in distillated water will 
deplete them of KCl. Please refill KCl-electrolyte (saturated or 3 mol) in time  

 

Do not touch the glass-diaphragm!

 Damages on the surface and attrition affect the 

performance negatively and might even destroy the electrode. 

  Before usage perform a visual check of the pH-electrode. If there are air-bubbles in the pH-

glass-diaphragm or the outer reference-electrode you can get them out by shaking the 
electrode downward.  

Содержание GE 014

Страница 1: ...Z03 0 0X 6C 11 Operating Manual pH electrodes GE xxx Keep for later use www greisinger de Companies Brands of the GHM English...

Страница 2: ...uct for consulting in case of doubt 2 Safety 2 1 Intended Use The pH electrodes must be used only according to its intended purpose and under suitable conditions The electrodes must be operated with s...

Страница 3: ...h can be damaged or destroyed when coming into contact with this chemical Caustic effect protective equipment required This Symbol indicates dangers to all living beings that may result in death or ac...

Страница 4: ...bility cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using the device The electrodes contain 3 molar KCL GE103 1mol l...

Страница 5: ...specification sheet of chemicals 3 Consider any statutory provisions guidelines and guidelines of chemicals when disposing This is also for accidentally spilled chemicals dried residues soiled rags or...

Страница 6: ...he temperature avoid interference electrostatic charges etc keep plug contacts clean and dry prevent electrodes except special waterproof versions from extended immersion above the shaft calibrate the...

Страница 7: ...ystems 3 4 Service life The service life of electrodes is normally at least 8 to 10 months When cared for properly this can usually increase to more than 2 years The actual life will vary depending on...

Страница 8: ...to 30 C Below 5 C the electrode might be damaged because of freezing of the electrolyte We recommend to store the electrode vertically with the cable to the upper side The pH electrode should be arra...

Страница 9: ...should be avoided as this might damage the electrode permanently Therefore a chemical cleansing is preferable Impurities Cleaners General residue Mild detergent Inorganic coatings common fluids for g...

Страница 10: ...E 100 can be used Some different applications requires a special electrode GE 100 is a universal electrode with two ceramic diaphragms and liquid GE 101 is preferably used for small sample amounts It...

Страница 11: ...disposal 6 1 Reshipment All devices returned to the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or sensor may be a ri...

Страница 12: ...Pt1000 sensor pH 0 14 0 60 C 100 S cm max 6 bar 3mol l KCl gel BNC and 4mm banana 2 m pressure resistant up to 6 bar with PG13 5 thread 2x ceramic cylinder PSU approx 12x120 mm GE 120 Cinch injection...

Отзывы: