5 | Operation
G 1910-02
NOTE
If the
Expanded settings
menu is processed completely, the changes are saved auto-
matically.
STOR
appears in the display. However, it can be exited at any time by press-
ing the
Function key
for 2 seconds. The changes made up that point are also saved.
If the product is switched off without saving the configuration, the last save value is
reproduced on the next start-up of the product.
5.2.5
Configure parameters of the expanded settings
menu
The following representation shows the available parameters and various configura-
tion options.
–
The
Expanded settings
Call-up of the expanded settings
1.
Select the desired parameter you would like to configure.
2.
Adjust the desired configuration in the selected parameter with the
Up key
and
Down key
.
3.
The available configuration options are listed for each parameter in the following
representation.
Parameter
Values
Meaning
Adjustment
[AL
OFF
No adjustment
2.PT
2-point adjustment
1.PT
1-point adjustment
KARD
Basic sensor adjustment
Only 0 ppm or 0.000 % can be selected as a set-
point value for the adjustment, such as, for example
on nitrogen or 400 ppm or 0.040 % for adjustment
on clean ambient air.
Gradient adjustment specification
[.SL
350
..
10,000
0.035
..
1.000
Setpoint value in ppm or in % for
1.PT
or
2.PT
adjust-
ment
0
/
400
0.000
/
0.040
Setpoint value in % or in ppm for
KARD
adjustment
Description
Prerequisite
Instruction
Representation