5 | Operation
G 1910-02
5
Operation
5.1
Commissioning
5.1.1
Explanation
The product is switched on with the
On/Off button
. It may be necessary to configure
the product after switching on. See
A self-test is conducted after switching on. The following increasing CO
2
values
1000
,
2000
,
4000
and
8000
are shown in the display.
TEST
is shown in the secondary display.
If an alarm is active and is in one of these relevant areas, it is triggered.
If
----
appears in the display after the self-test, the sensor is not ready for measure-
ment.
If batteries are drained and the product is not used for an extended time, it can take up
to 30 seconds until the measurement starts. If no measurement is received by the
sensor within 30 seconds, the product issues an alarm.
–
The rechargeable batteries have been charged via the micro USB socket.
–
Press the
On/Off button
.
Information about the configuration of the product appears in the display.
TEST
Self-test
An automatic self-test is carried out during system
start-up. The bar indicator for the threshold value is
displayed and the alarm is tested if it has been acti-
vated.
POFF
Automatic shut-
off
Automatic shut-off activated. The product is
switched off if no buttons have been pressed after
the adjusted time
–
The product is now ready for measurement.
5.2
Configuration
5.2.1
Explanation
The following steps describe how to adapt the product for your purposes.
NOTE
There are various configuration parameters available depending on the product ver-
sion and configuration. These can vary depending on the product version and configu-
ration.
5.2.2
Opening the configuration menu
In order to configure the product, you must first open the
Configuration
menu. The
menu is opened as shown in the illustration.
1.
Press the
Function key
for 2 seconds to open the
Configuration
menu.
2.
(ONF
appears in the display. Release the function key.
Description
Prerequisite
Instruction
Outcome of an action
Description
Prerequisite
Instruction