12 / 30 

B-H90.0.01.DB2-2.1 

5 | Operation 

G 1910-02 

Operation 

5.1 

Commissioning 

5.1.1

 

Explanation 

 

The product is switched on with the 

On/Off button

. It may be necessary to configure 

the product after switching on. See 

Configuration [

 12]

. 

A self-test is conducted after switching on. The following increasing CO

2

 values 

1000

2000

4000

 and 

8000

 are shown in the display. 

TEST

 is shown in the secondary display. 

If an alarm is active and is in one of these relevant areas, it is triggered. 

If 

----

 appears in the display after the self-test, the sensor is not ready for measure-

ment. 

If batteries are drained and the product is not used for an extended time, it can take up 
to 30 seconds until the measurement starts. If no measurement is received by the 
sensor within 30 seconds, the product issues an alarm. 

 

  The rechargeable batteries have been charged via the micro USB socket.

 

 

  Press the 

On/Off button

.

 

 

Information about the configuration of the product appears in the display. 

 

TEST

 

Self-test 

An automatic self-test is carried out during system 
start-up. The bar indicator for the threshold value is 
displayed and the alarm is tested if it has been acti-
vated. 

POFF

 

Automatic shut-
off 

Automatic shut-off activated. The product is 
switched off if no buttons have been pressed after 
the adjusted time 

  The product is now ready for measurement.

 

5.2 

Configuration 

5.2.1

 

Explanation 

The following steps describe how to adapt the product for your purposes. 

 

 

 

 NOTE 

 

There are various configuration parameters available depending on the product ver-
sion and configuration. These can vary depending on the product version and configu-
ration. 

5.2.2

 

Opening the configuration menu 

 

In order to configure the product, you must first open the 

Configuration

 menu. The 

menu is opened as shown in the illustration. 

 

1.

 

Press the 

Function key

 for 2 seconds to open the 

Configuration

 menu. 

2.

 

(ONF

 appears in the display. Release the function key. 

Description 

Prerequisite 

Instruction 

Outcome of an action 

 

Description 

Prerequisite 
Instruction 

Содержание G 1910-02

Страница 1: ...Operating manual G 1910 02 Compact CO2 monitor with alarm EN...

Страница 2: ...ng elements 10 4 4 Connections 11 5 Operation 12 5 1 Commissioning 12 5 1 1 Explanation 12 5 2 Configuration 12 5 2 1 Explanation 12 5 2 2 Opening the configuration menu 12 5 2 3 Configuring parameter...

Страница 3: ...Table of contents B H90 0 01 DB2 2 1 3 30 12 Service 29 12 1 Manufacturer 29 12 2 Repairs processing 29 12 3 Sales offices 29...

Страница 4: ...ty and warranty of the manufacturer for damages and consequential dam ages are voided with misuse disregarding this document disregarding safety notices assignment of inadequately qualified technical...

Страница 5: ...tion or a configuration of the product Formula Some instructions include a formula for a general understanding of a configuration programming or a setting of the product Outcome of an action Result co...

Страница 6: ...fety of the product can only be guaranteed if applicable safety precautions and the device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury...

Страница 7: ...er to life and health 3 0 Normal exhalation concentration ele vated breathing and pulse rate 2 0 20 000 1 0 10 000 Possible shortness of breath 0 5 5000 TWA Maximum for working condi tions 0 1 0 2 100...

Страница 8: ...rding of the mean value weighted over 8 hours TWA or 15 minutes STEL Monitoring of air quality 2 5 Qualified personnel For commissioning operation and maintenance the relevant personnel must have adeq...

Страница 9: ...ed on switched off and configured and the measurements and parameters can be adjusted and held with the operating elements The product is equipped with an integrated opti cal carbon dioxide sensor In...

Страница 10: ...in display Measurement of the current CO2 value Auxiliary display Display of the mean value Bar graph Visualisation of the CO2 value 4 3 Operating elements On Off button Press briefly Switch on the pr...

Страница 11: ...01 DB2 2 1 11 30 Function key Press briefly Freeze measurement Hold Return to measurement display Call up next parameter Long press 2s Start menu configuration ONF appears in the dis play 4 4 Connecti...

Страница 12: ...about the configuration of the product appears in the display TEST Self test An automatic self test is carried out during system start up The bar indicator for the threshold value is displayed and th...

Страница 13: ...onfiguration the last save value is reproduced on the next start up of the product 5 2 3 Configuring parameters of the configuration menu The following representation shows the available parameters an...

Страница 14: ...alarm is trig gered when the value is exceeded When the pre alarm is triggered it can be muted for 5 minutes To do this press any key The display shows LR AL 1 AL 2 Depending on the setting of the par...

Страница 15: ...ct is switched off 1 Press and hold the Down button 2 Press the On Off button to switch on the product 3 Release the On Off button after 1 second and then the Down button in order to call up the Expan...

Страница 16: ...ee Call up of the expanded settings menu 15 1 Select the desired parameter you would like to configure 2 Adjust the desired configuration in the selected parameter with the Up key and Down key 3 The a...

Страница 17: ...ise use the expected value of the test gas according to the analysis certificate or display value of the reference device The changed value is saved and the Expanded settings menu is closed STOR appea...

Страница 18: ...ter an optical band pass filter is passed an IR sensor measures the level of IR light wherein the CO2 values in the optical path are displayed 6 1 2 Design 1 Gas inlet 2 Gas outlet 3 IR lamp 4 Optical...

Страница 19: ...eries 7 2 Battery 7 2 1 Charge status display If the battery status display blinks the batteries are depleted and must be recharged However the device will still operate for a certain length of time I...

Страница 20: ...ies and dispose of them at a suitable collec tion point NOTE Read the following handling instructions before replacing rechargeable batteries and follow them step by step If disregarded the product co...

Страница 21: ...e air can take place Ensure that the setting of the gradient compensation setpoint SL was entered correctly in the Expanded settings menu If reliable values can no longer be displayed we recommend per...

Страница 22: ...bar indicator blinks 4 You can change the value of the gradient compensation with the up key and down key Otherwise confirm the preadjusted value with the function key The 1 point adjustment is finish...

Страница 23: ...dioxide concentration at the sensor is sta ble before the adjustment is started 3 If a stable correct value is reached the display flashes briefly an acoustic signal sounds and the bar indicator blink...

Страница 24: ...depleted Rechargeable bat tery discharged Rechargeable bat tery defect Charge battery Replace rechargeable bat tery AL ERR 1 Zero point adjust ment defective Incorrectly meas ured CO2 concentra tion...

Страница 25: ...40 C Bring product to room tem perature and restart charg ing process SYS ERR System error Error in the product Sensor defect Switch product on off Replace rechargeable bat teries Send in for repair...

Страница 26: ...ice must not be disposed of with household waste Return it to us freight pre paid We will then arrange for the proper and environmentally friendly disposal Private end users in Germany have the possib...

Страница 27: ...Current supply 2 AA NiMH batteries included in delivery Current requirement Rechargeable battery life approx 50 mA approx 60 mA with lighting Life approx 24 hours with NiMH batteries without backlight...

Страница 28: ...gas 5000 ppm CO2 476699 GZ 19 Adjustment set for G 1910 MiniFlo valve for 12 l gas cylinders quick connect adapter 4 mm closing screw gas bezel 479767 GZ 20 Gas connection for G 1910 Quick connect ada...

Страница 29: ...in our service centre Monday to Thursday from 8 00 to 16 00 Friday from 8 00 to 13 00 GHM Messtechnik GmbH GHM GROUP Greisinger Hans Sachs Str 26 Service Centre 93128 Regenstauf GERMANY Tel 49 940293...

Страница 30: ...e 26000 26999 35000 36999 40000 69999 Email vertrieb west ghm messtechnik de Tel 49 2191 9672 0 Fax 49 2191 9672 40 Post code 70000 97999 Email vertrieb sued ghm messtechnik de Tel 49 9402 9383 52 Fax...

Отзывы: