Dissolved oxygen measuring device
Operation | 7
B-H86.0.21.DB2-1.0
21 / 30
5. The changes are saved after running through the entire
Configuration
menu.
STOR
appears in the display. The
Configuration
menu can be exited from any arbitrary
parameter by pressing and holding the
Function key
for 2 seconds. The changes
made up that point are saved.
Representation
Call up menu
Next parameter Change value
Save changes
Discard
changes
2s
Press: Single
step
Hold: Rapid
change
2s
2s
Outcome of an action
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
7.2.3 Configuring parameters of the configuration menu
Description
The following representation shows the available parameters and various configura-
tion options.
Prerequisite
– The
Configuration
menu is open. See
Opening the configuration menu [
}
p. 20]
.
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
and
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.
Representation
Parameter
Values
Meaning
Input
INP
SAT
%
Oxygen saturation in per cent
(ONC
mg/l
Oxygen concentration in mg/l
(ONC
ppm
Oxygen concentration in ppm
Pressure
SET.P
500 .. 4000
Environmental pressure in hPa, corresponding to
mbar
Salinity correction
SAL
0 .. 70
Salinity in the measuring medium in PSU, corres-
ponding to g/kg