Dissolved oxygen measuring device

Error and system messages | 8

B-H86.0.21.DB2-1.0

25 / 30

8 Error and system messages

Display

Meaning

Possible

 

causes

Remedy

No display,
unclear char-
acters or no
response
when but-
tons are
pressed

Battery depleted

System error

Product is defective

Battery depleted

Error in the product

Product is defective

Replace battery

Send in for repair

BAT

Battery depleted

Battery depleted

Replace battery

BAT LO

Battery depleted

Battery depleted

Replace battery

(AL ERR.2

Slope is too low

Incorrect oxygen ref-
erence

Electrode contamin-
ated or defective

Perform calibration in damp
environmental air

Maintain the electrode

(AL ERR.3

Slope is too high

Incorrect oxygen ref-
erence

Electrode contamin-
ated or defective

Perform calibration in damp
environmental air

Maintain the electrode

(AL ERR.4

Incorrect calibration
temperature

Temperature too low
or too high

Range of 5..40 °C

(AL ERR.5

Time exceeded dur-
ing automatic calib-
ration

Unstable electrode
signal

Contaminated elec-
trode

Temperature not
equalised

Use calibration container

Maintain the electrode

Restart calibration

ERR.1

Measuring range ex-
ceeded

Measurement too
high

Electrode or product
defect

Faulty calibration

The measurement is above
the permissible range

Check electrode

Perform calibration

Send in for repair

ERR.2

Measuring range is
undercut

Measurement too
low

Electrode or product
defect

Check electrode

Send in for repair

SYS ERR

System error

Error in the product

Switch product on/off

Replace batteries

Send in for repair

Содержание G 1610

Страница 1: ...Operating manual G 1610 Dissolved oxygen measuring device Permanently connected oxygen sensor Waterproof EN...

Страница 2: ...10 4 3 Operating elements 10 5 Bases for measurement 12 5 1 Oxygen sensor 12 5 1 1 Explanation 12 5 1 2 Design 13 5 1 3 Service life 13 5 1 4 Operating position 14 5 1 5 Measurement accuracy 14 5 1 6...

Страница 3: ...ng parameters of the configuration menu 21 7 2 4 Adjustment of the measuring input 22 7 2 5 Configuring parameters of the adjustment menu 23 8 Error and system messages 25 9 Disposal 26 10 Technical d...

Страница 4: ...product 1 3 Legal notices The liability and warranty of the manufacturer for damages and consequential dam ages are voided with misuse disregarding this operating manual disregarding safety notices a...

Страница 5: ...product Outcome of an action Result consequence or effect of an instruction Emphases In order to simplify legibility and provide a clearer overview various sections informa tion are emphasised 1234 Di...

Страница 6: ...e NOTE This symbol indicates processes which can have a direct influence on operation or can trigger an unforeseen reaction in case of non observance NOTE This symbol instructs the use of eye protecti...

Страница 7: ...ct occur take the following measures Eyes Flush with flowing water for at least 15 minutes seek medical attention Skin Wash with large amounts of water for several minutes Clothing Remove immediately...

Страница 8: ...5 Qualified personnel For commissioning operation and maintenance the relevant personnel must have ad equate knowledge of the measuring process and use of the measurements for which purpose this docum...

Страница 9: ...he dust proof and waterproof design in accord ance with IP 65 67 and the 3 line illuminated display which offers overhead display at the push of a button The product can be switched on switched off an...

Страница 10: ...played O2 value with unit Bar graph Progress for calibration and visualisation of the electrode evaluation NOTE The unit display shows a rotating circle segment in the first position as long as the me...

Страница 11: ...B H86 0 21 DB2 1 0 11 30 Function key Press briefly Freeze measurement Return to measurement display Call up next parameter Long press 2s Start menu configuration ONF appears in the dis play Long pres...

Страница 12: ...ide The electrode contains potassium hydroxide This causes burns All contact with the skin clothing and eyes should be avoided Nevertheless should contact occur take the following measures Eyes Flush...

Страница 13: ...ubble ring on the membrane This can also be the cause if a sensor can no longer be calibrated The GWOK 02 membrane head is a spare part and can be re ordered sep arately Refill opening Electrolyte mus...

Страница 14: ...f air in the sensor This is usually due to incomplete filling or a leak due to improper fitting of the cap fill screw or membrane leak 5 2 Instructions for oxygen measurement The following must be obs...

Страница 15: ...f measuring accuracy specifying the current air pressure in the region based on meteorological data or the standard pressure based on sea level is suffi cient For example 0 m above NN 1013 hPa 300 m a...

Страница 16: ...fill opening is filled up to the rim with KOH screw the in the seal screw again 5 Rinse off excess KOH and screw the sensor on with the membrane head In the process if there are any recognisable air b...

Страница 17: ...depleted and must be replaced However the device will still operate for a certain length of time If the BAT display text appears in the main display the battery voltage is no longer ad equate for ope...

Страница 18: ...hed on NOTE Calibration takes place in air saturated with water vapour The GCAL 3610 calibration container or the storage bottle can be used for this purpose The membrane of the electrode must be dry...

Страница 19: ...ng the Function key The product restarts and the value of the last suc cessful calibration is restored 6 3 2 Battery indicator If the empty frame in the battery display blinks the batteries are deplet...

Страница 20: ...the electrode prior to starting the measurement See Calibration and adjustment service 7 2 Configuration 7 2 1 Explanation The following steps describe how to adapt the product for your purposes NOTE...

Страница 21: ...start up of the product 7 2 3 Configuring parameters of the configuration menu Description The following representation shows the available parameters and various configura tion options Prerequisite T...

Страница 22: ...ion is closed if no button is pressed for 2 minutes Any changes made up to that point are not saved C END appears in the display 7 2 4 Adjustment of the measuring input Description The temperature inp...

Страница 23: ...he Adjustment menu is open See Adjustment of the measuring input p 22 Instruction 1 Select the desired parameter you would like to configure 2 Adjust the desired configuration in the selected paramete...

Страница 24: ...zero point correction setpoint zero point T OF 0 2 C 0 0 C 0 2 C T SL setpoint gradient correction display gradient correction T OF 1 100 T SL 37 0 C 36 6 C 0 2 1 100 0 54 Outcome of an action The cha...

Страница 25: ...is too high Incorrect oxygen ref erence Electrode contamin ated or defective Perform calibration in damp environmental air Maintain the electrode AL ERR 4 Incorrect calibration temperature Temperatur...

Страница 26: ...information base online at www ghm group de and sent it in with the product NOTE The device must not be disposed of with household waste If the product is disposed of please take it to a municipal col...

Страница 27: ...ode Connections Permanently connected oxygen sensor Operating conditions Device 20 to 50 C 0 to 95 r h temporarily 100 r h Electrode 0 40 C Storage temperature 0 40 C Current sup ply 2 AA battery incl...

Страница 28: ...4 GSKA 3600 PVC protective cap submerged for use in still bodies of water 607267 GSKA 3610 Red brass protective cap salt water resistant also suit able for use with greater depths or with a flow 61137...

Страница 29: ...de www greisinger de WEEE reg no DE 93889386 12 2 Repairs Defective products are repaired professionally and quickly in our service centre Open hours and contact Monday to Thursday from 8 00 to 16 00...

Страница 30: ...rance GHM GROUP France SAS Parc des Pivolles 9 Rue de Catalogne 69150 D cines Charpieu Lyon FRANCE Phone 33 4 72 37 45 30 a jouanilou ghm group fr India GHM Messtechnik India Pvt Ltd 209 Udyog Bhavan...

Отзывы: