22
Accuracy
(cont’d)
Frequency—Line Level
Range: 5.000 Hz to 200.00 kHz
Accuracy: ± (0.02% + 4 digits)
AC Function
Range
Sensitivity
(Sine RMS)
Range
AC Function
Range
Sensitivity
(Sine RMS)
Range
500 mV
100 mV
10 Hz to 200 kHz
500 µA
50 µA
10 Hz to 10 kHz
5 V
0.5 V
5000 µA
500 µA
50 V
5 V
10 Hz to 100 kHz
50 mA
5 mA
500 V
50 V
500 mA
50 mA
1000 V
500 V
10 Hz to 10 kHz
5 A
1 A
10 Hz to 3 kHz
VFD 5 V
0.5 V to 2 V*
10 Hz to 440 Hz
10 A
10 A
VFD 50 V
5 V to 20 V*
*
VFD sensitivity linearly decreases from 10% of
range @ 200 Hz to 40% of range @ 440 Hz.
VFD 500 V 50 V to 200 V*
Frequency —Logic Level
Range: 5.000 Hz to 2.0000 MHz
Accuracy: ± (0.002% + 4 digits)
Sensitivity: 2.5 Vp square wave
% Duty Cycle
Range: 0.1% to 99.99%
Accuracy: ± (3 digits/kHz + 2 digits)
Input Frequency: 5 Hz to 500 kHz, 5 V logic family
Diode Test
Measuring Range: 2.0000 V
Test Current (typical): 0.4 mA
Open Circuit Voltage: < 3.0 VDC
Accuracy: ± (1% + 0.0001 V)
Continuity
Tone Threshold: Between 20 Ω and 200 Ω
Response Time: < 100 µs
dBm
The range and accuracy depend on the AC voltage function used and the selected reference
impedance.
Selectable Reference Impedance: 4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500,
600, 800, 900, 1000, 1200 Ω
Crest Capture (Voltage and Current) for Crests > 0.8 ms in duration
Accuracy: Specified accuracy ± 100 digits
Resolution: 5000 digits
Содержание DM-860A
Страница 14: ...14 Typical Measurements Voltage Measurement Current Measurement ...
Страница 16: ...16 Typical Measurements Temperature ...
Страница 26: ...26 ...
Страница 39: ...DM 860A 39 Mediciones más comunes Medición de voltaje Medición de corriente ...
Страница 41: ...DM 860A 41 Mediciones más comunes Temperatura ...
Страница 63: ...DM 860A 63 Mesures types Mesure de tension Mesure d intensité ...
Страница 65: ...DM 860A 65 Mesures types Température ...
Страница 75: ...DM 860A 75 ...