CMI-100
15
Peak Hold Accuracy Table
Value
Range
Accuracy
Frequency
Input
Range
Impedance
3.999 V
±
(2.5% + 0.01 V)
50 to 60 Hz
≈
11 M
Ω
AC Voltage
39.99 V
±
(2.5% + 0.1 V)
(Manual Range)
399.9 V
±
(2.5% + 1 V)
50 to 60 Hz
≈
10 M
Ω
600 V
±
(2.5% + 10 V)
AC Current
399.9 A
±
(2.5% + 1 A)
50 to 60 Hz
(Manual Range)
1000 A
±
(2.5% + 10 A)
Diode Test
Test Current (Typical): 0.8 mA
Open Circuit Voltage: 3 V maximum
Accuracy:
±
(1% + .002 V)
Continuity
Threshold: Tone indicates circuit resistance is
≤
40
Ω
.
Open Circuit Voltage:
≤
0.5 V
Additional Tolerance Due to Crest Factor
Crest Factor
Error (% of reading)
1 - 3
0.5%
3 - 5
3%
5 - 7
6%
Accuracy