G36p
Oxygen Analyzer
Version 2.11 – Revision January 2021 – Part 01381
27
H&L alarm manual res
is used to switch the manual reset function of the High or Low
alarm ON/OFF.
When the “H&L alarm manual res” is OFF, the High or Low alarm will disappear auto-
matically when the reading of the O
2
value is no longer in the alarm zone (O
2
level
alarms).
When the “H&L alarm manual res” is ON, it is required a deliberate manual action to re-
set either High or Low alarm.
To reset an alarm, enter
Tools menu
and choose
Alarm manual reset
.
HH alarm manual res
is used to switch the manual reset function of the High High
alarm ON/OFF.
When the “HH alarm manual res” is OFF, the High High alarm will disappear automati-
cally when the reading of the O
2
value is no longer in the alarm zone (O
2
level alarms).
When the “HH alarm manual res” is ON, it is required a deliberate manual action to reset
the High High alarm.
To reset an alarm, enter
Tools menu
and choose
Alarm manual reset
.
Note
If the analyzer is intended to detect a potentially hazardous gas concentration and the
Analyzer alarms is set ON, the manual reset functions of the intended alarm should
also be set ON in order to compliance with the common safety requirement.
5.3
Setting of the Analog Input
The route for setting the analog input is
Tools menu\Service (Password level 1:
9490)\Analog in/out settings\Analog input 1
.
Both analog inputs are not used by default. Therefore, the parameters in both analog in-
put settings must not be edited.
However, if the analyzer has been specifically ordered with preset analog inputs, please
see Appendix 74H10.1 for more information.