Analyzer Installation & Connections
10
3.3
Safety Aspects
Warning
Follow the operating instructions!
Make sure that all power and signal cables are connected correctly before operating
the analyzer.
The sensor must not be exposed to strong mechanical shocks. Otherwise the sensor
element may crack without visible damage.
The sensor and sensor housing are hot and can cause severe burning of personnel if
not handled with care.
The G36p Oxygen Analyzer is only suitable for installation in a safe, non-hazardous
area and not suitable for use with flammable sample gases.
The installation and operation of the G36p
Oxygen Analyzer and associated equip-
ment must be carried out by skilled, trained, and certified personnel. Green Instru-
ments A/S does not take any responsibility for the operation of the analyzer and as-
sociated equipment whatsoever.
The analyzer must only be applied as described in this instruction manual. If the ana-
lyzer is used in a manner not specified by Green Instruments, the protection provid-
ed by the analyzer may be impaired.
Successful and safe operation of this equipment is dependent upon proper handling,
installation, operation, and maintenance.
Hazardous voltage!
Always disconnect the power before installing or servicing the analyzer. Ignoring this
warning can result in severe personal injury or material damage. Read the manual care-
fully to ensure that all power and signal leads are connected correctly.
Make sure that the analyzer is connected to the correct voltage (see rating marked on the
analyzer name plate).
Circuit breaker!
The installation must include a way to switch off the electrical power by a clearly marked
switch or circuit breaker external to the analyzer. The external switch or circuit breaker
shall be placed in close proximity to the analyzer and within easy reach of the operator.
Overload protection!
For compliance with the IEC 61010-1 (2003) safety requirements, the installation must
include a means of overcurrent protection to provide protection against excessive energy
being drawn from the power supply system in case of a fault in the analyzer.