@SWB7 USER MANUAL
12
1. Press the key to display the G value of manufacture place.
2. If set the G value of local press the key and then press the or
key and key to input the new G value.
3. Press the key to confirm.
Weight Calibrations
(
*
)
1. In weight mode, press the and keys, the display will show
ECF-1
.
2. Press the or keys to select
ECF-1
,
ECF-2
or
ECF-3
ECF-1
Zero and Span Calibration
1. Press the key to enter, display will show
cALZ
.
2. Press the key to calibrate zero point.
3. The display will show the calibration weight
00.6.000
.
4. Use the keys and to select the digit,press key to input the
weight value.
5. Put the calibration weight on the platter and press the key to calibrate.
The scale will return to weighing mode automatically.
ECF-2
Zero Calibration
1. Press the key to enter, display will show
cALZ
.
2. Press the key to calibrate zero point.
3. The scale will return to weighing mode automatically.
ECF-3
Span Calibration
1. Press the key to enter, display will show the calibration weight
00.6.000
.
2. Use the keys and to select the digit,press key to input the
weight value.
3. Put the calibration weight on the platter and press the key to calibrate.
The scale will return to weighing mode automatically.